![vacuubrand ME 16C NT-ME 16C NT+EK Instructions For Use Manual Download Page 24](http://html1.mh-extra.com/html/vacuubrand/me-16c-nt-me-16c-nt-ek/me-16c-nt-me-16c-nt-ek_instructions-for-use-manual_3202245024.webp)
page 24 of 88
Technical data
General technical data valid for all pumps
ATEX approval if the ATEX marking
is shown on the rating plate
Inner part (pumped gases)
II 3/- G Ex h IIC T3 Gc X
Internal Atm. only
Tech.File: VAC-EX02
Maximum permissible inlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum permissible outlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum pressure difference between
inlet and outlet
psi
(bar)
16
(1.1)
Maximum permissible pressure (abso-
lute) at gas ballast valve
psi
(bar)
17.5
(1.2)
Permissible ambient temperature
storage / operation
°F
(°C)
14 to 140 / 50 to 104
(-10 to +60 / +10 to +40)
Permissible relative atmospheric moisture
during operation (no condensation)
%
30 to 85
Maximum permissible installation altitude
above mean sea level
ft
(m)
6500
(2000)
Rated motor power
hp
(kW)
0.71
(0.53)
No-load speed
rpm
1650
Maximum permissible range
of supply voltage (±10%)
Attention: Observe specifications of rating plate!
100-120 V~ 50/60 Hz
200-230 V~ 50/60 Hz
Maximum rated current at:
100-120 V~ 50/60 Hz
200-230 V~ 50/60 Hz
A
A
8.0
3.5
Device fuse
2 slow blow fuses
250 V / 8AT - 5x20
Motor protection
temperature sensor in the
winding
Overvoltage category
II
Degree of protection IEC 60529
IP 40
Pollution degree
2
Inlet
small flange KF DN 25
Summary of Contents for ME 16C NT-ME 16C NT+EK
Page 29: ...page 29 of 88 ME 16C NT MD 12C NT MV 10C NT 1 5 3 2 6 4 8 7 9...
Page 49: ...page 49 of 88 Disassembling the pump housing 1 3 2...
Page 50: ...page 50 of 88 4 6 5 1 2 4 4x 2 5 mm 2 5 mm...
Page 51: ...page 51 of 88 7 9 8 4 4x without EK...
Page 52: ...page 52 of 88 10 12 11 4 2x 4 4x Don t loosen this screw...
Page 53: ...page 53 of 88 13 15 14 5 2x 5 2x...
Page 54: ...page 54 of 88 16 18 17 2 5 mm...
Page 56: ...page 56 of 88 Replacing the diaphragms 19 21 20 Clean 5 8x 20696867 20696839...
Page 57: ...page 57 of 88 22 24 23 1 2 3 Pay attention to number of washers...
Page 58: ...page 58 of 88 25 27 26...
Page 61: ...page 61 of 88 34 36 35 Clean 20696867 20696839...
Page 62: ...page 62 of 88 37 39 38 IN IN EX EX 4x...
Page 63: ...page 63 of 88 40 42 41 IN EX T20 4x 2 2 ft lbf 3 Nm...
Page 64: ...page 64 of 88 43 45A 44 Maintain all pump heads in the same way MD 12C NT...
Page 65: ...page 65 of 88 45B 45C 45D ME 16C NT MV 10C NT ME 16C NT MD 12C NT MV 10C NT...
Page 67: ...page 67 of 88 4A 3 2 size 2 2x Version A valve 24 4 mm...
Page 68: ...page 68 of 88 5A 20696839 24 4 mm 26 mm 4B 5B 20696839 37 mm 40 mm Version B valve 37 mm...
Page 69: ...page 69 of 88 6 7 size 2 2x 7A without EK size 2 2x...
Page 71: ...page 71 of 88 Assembling the pump housing 46 47 48 5 2x 8 9 ft lbf 12 Nm...
Page 72: ...page 72 of 88 50 49 5 2x 8 9 ft lbf 12 Nm 51...
Page 73: ...page 73 of 88 53 52 4 4x 3 7 ft lbf 5 Nm 54 4 2x 3 7 ft lbf 5 Nm...
Page 74: ...page 74 of 88 56 55 4 4x 3 7 ft lbf 5 Nm 57...
Page 75: ...page 75 of 88 59 58 60 4 4x 3 7 ft lbf 5 Nm...
Page 76: ...page 76 of 88 62 61 63...
Page 79: ...page 79 of 88 4 20638821 20638821 5 6...
Page 80: ...page 80 of 88 7 T20 4x size 1 4x 8...
Page 83: ...page 83 of 88 Declarations and certificates...
Page 84: ...page 84 of 88...
Page 87: ...page 87 of 88...