page 17 of 88
Safety during operation
➨
Adopt suitable measures to prevent the release of dan-
gerous, toxic, explosive, corrosive, noxious or pollut-
ing fluids, vapors and gases. To prevent any emission
of such substances from the pump outlet, install an
appropriate collecting and disposal system and take
protective action for pump and environment.
➨
You must take suitable precautions to prevent any for-
mation of explosive mixtures in the pump chamber or
at the outlet of the pump. In case, e.g., of a diaphragm
failure, mechanically generated sparks, hot surfaces
or static electricity may ignite these mixtures. Use inert
gas for gas ballast or venting, if necessary.
➨
Drain appropriately or otherwise remove any poten-
tially explosive mixtures at the outlet of the pump, or
dilute them to non-explosive concentrations.
➨
Never operate this pump if it has a damaged cord or
plug.
+
If the pump is not working properly, has been dropped or
has fallen into water, contact your pump service provider.
+
Prevent any part of the human body from coming into
contact with vacuum.
+
If there is an exhaust isolation valve, make sure that you
cannot operate the equipment with the valve closed to
avoid a
risk of bursting!
+
Always provide a free and pressureless exhaust outlet
to avoid damage to pump valves and risk of bursting.
+
Check the overpressure safety relief device at the ex-
haust waste vapor condenser at appropriate intervals.
+
Comply with applicable regulations when disposing of
chemicals. Take into consideration that chemicals may
be contaminated. Take adequate precautions to pro-
tect people from the effects of dangerous substances
Summary of Contents for ME 16C NT-ME 16C NT+EK
Page 29: ...page 29 of 88 ME 16C NT MD 12C NT MV 10C NT 1 5 3 2 6 4 8 7 9...
Page 49: ...page 49 of 88 Disassembling the pump housing 1 3 2...
Page 50: ...page 50 of 88 4 6 5 1 2 4 4x 2 5 mm 2 5 mm...
Page 51: ...page 51 of 88 7 9 8 4 4x without EK...
Page 52: ...page 52 of 88 10 12 11 4 2x 4 4x Don t loosen this screw...
Page 53: ...page 53 of 88 13 15 14 5 2x 5 2x...
Page 54: ...page 54 of 88 16 18 17 2 5 mm...
Page 56: ...page 56 of 88 Replacing the diaphragms 19 21 20 Clean 5 8x 20696867 20696839...
Page 57: ...page 57 of 88 22 24 23 1 2 3 Pay attention to number of washers...
Page 58: ...page 58 of 88 25 27 26...
Page 61: ...page 61 of 88 34 36 35 Clean 20696867 20696839...
Page 62: ...page 62 of 88 37 39 38 IN IN EX EX 4x...
Page 63: ...page 63 of 88 40 42 41 IN EX T20 4x 2 2 ft lbf 3 Nm...
Page 64: ...page 64 of 88 43 45A 44 Maintain all pump heads in the same way MD 12C NT...
Page 65: ...page 65 of 88 45B 45C 45D ME 16C NT MV 10C NT ME 16C NT MD 12C NT MV 10C NT...
Page 67: ...page 67 of 88 4A 3 2 size 2 2x Version A valve 24 4 mm...
Page 68: ...page 68 of 88 5A 20696839 24 4 mm 26 mm 4B 5B 20696839 37 mm 40 mm Version B valve 37 mm...
Page 69: ...page 69 of 88 6 7 size 2 2x 7A without EK size 2 2x...
Page 71: ...page 71 of 88 Assembling the pump housing 46 47 48 5 2x 8 9 ft lbf 12 Nm...
Page 72: ...page 72 of 88 50 49 5 2x 8 9 ft lbf 12 Nm 51...
Page 73: ...page 73 of 88 53 52 4 4x 3 7 ft lbf 5 Nm 54 4 2x 3 7 ft lbf 5 Nm...
Page 74: ...page 74 of 88 56 55 4 4x 3 7 ft lbf 5 Nm 57...
Page 75: ...page 75 of 88 59 58 60 4 4x 3 7 ft lbf 5 Nm...
Page 76: ...page 76 of 88 62 61 63...
Page 79: ...page 79 of 88 4 20638821 20638821 5 6...
Page 80: ...page 80 of 88 7 T20 4x size 1 4x 8...
Page 83: ...page 83 of 88 Declarations and certificates...
Page 84: ...page 84 of 88...
Page 87: ...page 87 of 88...