6. TROUBLE SHOOTING
Syndrome
Scale cannot be turned on
Check:
Is the scale charged?
Action:
Recharge the scale for at least 8 hours
before first time use or plug in the power
adaptor before power on the scale.
Check
Is the power adaptor inserted properly into
both the electricity outlet and the DC inlet
of scale.
Action:
Secure both ends of the power adaptor and
try again.
Syndrome
Scale turned off automatically
Check
Is
Auto Power Off Function
employed?
Action
Refer to
4.2.2
to disable
Auto Power Off
Function
.
Check
Is the
LO-BAT INDICATOR
on?
Action
Apply the power adaptor.
Syndrome
Rated capacity cannot be reached
Check
Is the
NET INDICATOR
on?
Action:
Turn the scale off. Remove all loads from
platter and turn on again.