10. NET INDICATOR 

This indicator appears to indicate the tare function 

is in operation and weight displayed is net weight 

 

11. LO-BAT INDICATOR 

When  this  indicator  appears,  recharge  the  scale 

immediately. Failure to do so may cause unrecoverable 

damage to the rechargeable battery inside scale 

 

12.

 

STANDBY INDICATOR  

This indicator appears to indicate scale is in standby 

mode (power saving function) 

 

13. IN-CHARGE INDICATOR 

Red color: Recharging battery 

Green color: Charging completed 

 

 

 

 

 

Summary of Contents for ADW-1500E

Page 1: ......

Page 2: ...CONTENTS CAUTIONS 1 INSTALLATION 2 SPECIFICATIONS 3 KEYBOARD LAYOUT AND DESCRIPTION 4 INITIAL SETUP 5 INSTRUCTION FOR USE 6 TROUBLE SHOOTING 7 DAILY CARE AND MAINTENANCE ...

Page 3: ...ult do not use this scale in where or when the environment condition falls beyond as those listed on SPECIFICATIONS 3 Do not attempt to open this unit or conduct any trouble shootings other than those listed on TROUBLE SHOOTING No user serviceable parts inside 4 Do not submerge scale into water 5 Do not use pressurized water nozzle for cleaning 6 Do not use this scale for chemical substances or in...

Page 4: ...8 Always apply the battery cap when not in charging 9 Check desiccant underneath the scale periodically Should any part of the desiccant turns to pink color contact your dealer for scale inspection ...

Page 5: ...and make sure that the following accessories are included This operation manual Scale x 1 Plastic platter x 1 Stainless steel cover x 1 Screw x 1 Rubber cap x 1 Power adaptor x 1 Contact your dealer if any item is missing FIG 1 ...

Page 6: ...r cap FIG 1 C 1 3 Place the stainless steel platter FIG 1 D onto the plastic platter 1 4 Power on the scale a Before charging the scale or power on with the adaptor supplied Please make sure the input voltage listed on the back of the adaptor matches with the local voltage If not do not use it and contact your dealer immediately ...

Page 7: ...e first time use charge the scale for at least 8 hours to ensure the best performance c After battery is fully charged remove adaptor from scale Apply the battery cap FIG 2 E to seal the DC input jack of scale ...

Page 8: ... Zero Range 2 Max Min Load 20e Display LED Display Power Source By Built in Rechargeable Battery or AD DC Power Adaptor Operating Environment 0o 40o C 32o 104o F Non condensed R H 85 Specifications subject to change without notice The minimum load of 15kg for Non OIML application is 25e NOTE 1 OIML Legal for trade 1 3000 Non OIML Not legal for trade 1 6000 1 7500 NOTE 2 Display resolution of avoir...

Page 9: ...ERO KEY Press this key to a turn scale on or reset display value to zero b move cursor to the right in function setup 3 MODE KEY Press this key to a shift among different weight units b increase value in function setup 4 TARE KEY Press this key to a tare off the weight of a container ...

Page 10: ...t check function 7 HI INDICATOR This indicator appears to indicate the weight detected above the high limit set Weight check function 8 AVOIRDUPOIS WEIGHT UNITS INDICATOR This indicator appears to indicate the weight being is avoirdupois weight units lb Two avoirdupois weight units are available i e decimal pound e g 5 62 means 5 62 lb and pound ounce e g 2 15 01 means 2 lb and 15 1 ounce 9 ZERO I...

Page 11: ...ATOR When this indicator appears recharge the scale immediately Failure to do so may cause unrecoverable damage to the rechargeable battery inside scale 12 STANDBY INDICATOR This indicator appears to indicate scale is in standby mode power saving function 13 IN CHARGE INDICATOR Red color Recharging battery Green color Charging completed ...

Page 12: ... for leveling condition 4 2 INTERNAL SETTING 4 2 1 Display Segment Check User can enter this function to check to see if the display and backlight if purchased are functioning properly a Scale is off b Press and hold TARE then press ON ZERO c Scale displays F 1 d Press TARE once e Scale displays F 2 f Press MODE and all segments will light up to allow user to check the condition of display g Press...

Page 13: ...ARE three times e Scale displays F 4 f Press MODE to shift between 0 _OFF and 4 _OFF To enable Auto Power Off Function press TARE when 4 _OFF appears To disable Auto Power Off Function press TARE when 0 _OFF appears g Press ON ZERO to restart the scale or TARE for next function setup 4 2 3 Select RS 232 Baud Rate Protocol This scale is equipped with 4 RS232C output baud rate and 2 transmission pro...

Page 14: ...preferred transmission protocol shows on display press ON ZERO to confirm and restart the scale or TARE for next function setup NOTE The default data output format of the scale is Baud Rate 9600 Data Bit 8 Stop Bit 1 Parity None 4 2 4 Select Data Transmission Mode This scale provides 2 types of transmission mode depending on the requirement and environment a Scale is off b Press and hold TARE then...

Page 15: ... is stabled trAn2 is to transmit single data only after reading is stabled However scale will not send any data if weight detected is less than the minimum load 4 2 5 Select Auto Tare Function This scale is equipped with auto tare function which tares off the initial weight automatically when this function is engaged The tare weight is automatically cleared when everything is removed from platter ...

Page 16: ...to select between Tr _on or Tr oFF To enable Auto Tare Function press TARE when Tr _on appears To disable Auto Tare Function press TARE when Tr oFF appears g Press ON ZERO to confirm and restart the scale or TARE for next function setup ...

Page 17: ... surface c The scale is turned on d The ZERO INDICATOR is on If not press ON ZERO to set display to zero 5 2 WEIGHING a Always place an object onto the platter gently Excessive force applied to platter may cause damages to sensor inside scale b The weight of the object would be displayed automatically c It is a good practice to remove all loads from platter after weighing It would prolong the life...

Page 18: ...dealer for more information 5 4 SMART POWER SAVING FUNCTION When this function is enable the displays digits will be turned off when a stable weight is detected and remains stable for about 10 seconds Standby indicator appears to indicate scale is now in standby mode Display will be automatically turned on again when a different weight is detected 5 4 1 Select Power Saving Setting The scale is equ...

Page 19: ... negative value when scale is not at its zero After ON ZERO is pressed the ZERO INDICATOR appears Refer to SPECIFICATIONS for maximum zero range NOTE When scale is set at OIML and application is legal for trade User only needs to press ON ZERO once when setting the value to zero The re zero function will take place when reading is stabled When scale is set at Non OIML and application is not legal ...

Page 20: ... is pressed the NET INDICATOR would appear Refer to 5 2 for weighing procedures Weight displayed under is the net weight of the subject matter Refer to SPECIFICATIONS for maximum tare range NOTE When scale is set at OIML and application is legal for trade User only needs to press TARE once when tare off a weight The tare function will take place when reading is stabled When scale is set at Non OIM...

Page 21: ...d with Weight Check Function An audio and video signal is generated to reflect the weight check result 5 8 1 Setting the High Low Weight Limits and Buzzer Warning a Remove all loads from platter b Turn scale on c Make sure that ZERO INDICATOR is on if not press ON ZERO d Press and hold ON ZERO then press TARE e The HI INDICATOR appears and the cursor is flashing scale is now in upper weight limit ...

Page 22: ... warning parameter setup l Press MODE until the desired buzzer output parameter appears NOTE bb_L buzzer warning when weight detected below the low weight limit set bb_no no buzzer warning bb_S buzzer warning when weight detected within the high and low weight limit set bb_H L buzzer warning when weight detected below above the low high weight limit set bb_H buzzer warning when weight detected abo...

Page 23: ... high weight limit is less than the value of the weight limit set To disable Weight Check Function simply set the high weight limit to zero or set the buzzer warning to bb_no to turn off the warning 5 8 2 Weight Check Transaction Place a subject matter onto the platter The Weight Check result is displayed on the display a The HI INDICATOR will appear when weight detected over the high weight limit...

Page 24: ...lay will flash the extended digit for 5 seconds before return to normal 5 11 RECHARGE THE SCALE The scale is equipped with a built in rechargeable battery When the LO BAT INDICATOR appears recharge the scale immediately Failure to do so may cause unrecoverable damage to the rechargeable battery Charging is completed when the color of the IN CHARGE INDICATOR turns green ...

Page 25: ...th the electricity outlet and the DC inlet of scale Action Secure both ends of the power adaptor and try again Syndrome Scale turned off automatically Check Is Auto Power Off Function employed Action Refer to 4 2 2 to disable Auto Power Off Function Check Is the LO BAT INDICATOR on Action Apply the power adaptor Syndrome Rated capacity cannot be reached Check Is the NET INDICATOR on Action Turn th...

Page 26: ...n or RFI Action Place the scale away from all disturbances Check Is the scale calibrated correctly Action Contact your dealer Check Is the correct weight unit employed Action Refer to 5 3 for weight unit setting and usage Syndrome Display shows 00000 after counting down Check Is the scale overloaded Action Remove some loads from platter Check Is the stainless steel platter cover placed on the plat...

Page 27: ...cally Re calibrate the scale if necessary NOTE In some countries calibration is restricted to be done by an authorized qualified agent only Contact your dealer for more information 7 4 It is a good practice to apply the dust cover when operating the scale 7 5 The scale must be placed horizontally during transportation or long time storage 7 6 Remove platter from scale before transportation or long...

Page 28: ...he following EC directives correspond au modèle décrit dans le certificat d approbation CE de type aux exigences de la directive 90 384 CEE modifiée et aux exigences des directives CE suivantes Diese Erklärung gilt nur in Verbindung mit einer Konformitätsbescheinigung einer benanten Stelle This Declaration of conformity is only valid with a certificate of conformity issued by a notified body Cette...

Reviews: