minimum size is thus 4. The WIN commands must specify start addresses lower or equal to
corresponding stop addresses and must be within the allowed range for the supported FPA for these
commands to complete without error.
A smaller column number may be set, even as there is no speed improvement. This will allow the user to
save smaller images and video sets.
5.20.1.
Set Window Column Start Number
Description
Sets the start column number for the ROI. The value must be an
EVEN number and compatible with the current column stop
number for the command to complete without error.
Setting Type
Operational
Command WIN:COL:START
value
Parameters
value
Return Values
none
Range
0 to 636
Type unsigned
integer
Example
WIN:COL:START 10 -- start read at col 10
5.20.2.
Get Window Column Start Number
Description
Returns the start column number for the ROI.
Setting Type
Operational
Command WIN:COL:START?
Parameters none
Return Values
value
Range
0 to 636
Type unsigned
integer
Example
WIN:COL:START?
-- query command
420
--
return
value
5.20.3.
Set Window Column Stop Number
Description
Sets the stop column number for the ROI. The value must be ODD
and compatible with the current column start number for the
command to complete without error
Setting Type
Operational