5.12.1.
Set Trigger Mode
Description
Sets the trigger and timing modes. Note: Execution of this
command that results in a change in trigger mode will also apply a
reset to the trigger and FPA scan digital logic clearing any existing
trigger or scan errors.
Setting Type
Global
Command TRIG:MODE
mode
Parameters
mode
Return Values
none
Range
0
Internally triggered, internally timed
1
Externally triggered, internally timed
2
Externally triggered, externally timed
3
Externally gated, internally timed
Type unsigned
integer
Example TRIG:MODE
1
5.12.2.
Get Trigger Mode
Description
Returns the trigger and timing mode.
Setting Type
Global
Figure 17. Timing sequence for trigger mode 3