Other
settings
One Measurement Only, when enabled, returns the counts for only one Measurement Time, and then
stops execution (otherwise measurements run continuously). This is useful, for example, when changing
settings between each measurement.
Time tagger Resolution reports the resolution of the Coincidence Logic Unit in nanoseconds.
Coinc. Window controls the time window in nanoseconds within which counts on different channels are
counted as coincident.
Actual Coinc. Window reports the coincidence window actually used in nanoseconds, which must be a
multiple of the Coincidence Logic Unit’s resolution.
Measurement Time chooses the time in seconds between reading the Coincidence Logic Unit’s buffer,
meaning Counts are displayed as summed over the Measurement Time.
Actual Measurement Time reports the real time in seconds between reading the Counts, due to finite
processing time of the PC.