w w w . u q d e v i c e s . c o m
L o g i c 1 6 U s e r M a n u a l
| 18
3.4
Firmware Update/Change
The following section outlines how to perform firmware updates and changes. For best results it is
important that the Logic16 Unit is running the latest FPGA firmware. The same procedure for updates is
used to switch from Configuration A to B, or vice versa. Tip: You can check the firmware version quickly
using Timetag Explorer
10
.
You need the following items:
Logic16 Unit
Small Phillips or flathead screw driver
JTAG link cable (e.g. JTAG-cable Xilinx, Xilinx parallel cable, HS2 Digilent)
Xilinx Lab Tools – iMPACT (Free download from http://www.xilinx.com/downloads)
o
Find and download latest ISE Design Tools available for your PC under Vivado
(HW Developer) in ISE Archive.
o
Follow setup instructions given by Xilinx and ensure Xilinx iMPACT is
accessible.
Firmware PROM files provided by UQD (found in the CD drive provided)
Time Tag Explorer (optional)
Perform the following procedures to update/change the Firmware:
i
Connect Logic16 to PC with a JTAG Cable
Modules manufactured after January 2015: connect JTAG Cable to the designated
JTAG connector found on the rear of the device.
Modules manufactured before January 2015: remove the right-hand side panel
(unscrew). Connect the JTAG Cable to the FPGA board and then to your computer.
ii
Connect device to a power supply .
Connect Logic16 to a DC power supply (12 V @ ≥1.5 A) or a NIM crate (12 V @ 1.0 A
and -12V @0.4 A). The ports are found on the rear side of the device.
10
Upon connection, it will appear in the window on the Time Tag Explorer’s TimeTag tab