w w w . u q d e v i c e s . c o m
L o g i c 1 6 U s e r M a n u a l
| 13
Figure 3-II
Time Tag Filtering:
Signals S1, S2 and S3 are input on
Logic16. With Time Tag Filter minimum group number set to 2 and
maximum time as shown, only pulses with their active edge within the
filter are tagged.
Gating
Edge Gating
The Edge Gate trigger is located on input 8. This gate opens for a user-specified time, after there
is an active edge on input 8. The gate width should be less than the period of the trigger signal or
the results will be unreliable. The position of the gate can be adjusted to compensate for cable
delays and similar effects. This can be a negative and positive time interval.
7
7
In configuration B, Logic16 input 8 does not time tag when using the external clock input (10
MHz). The edge gate can still be triggered on/off in this case with an input signal, however will
NOT time tag.