EX-MANUAL-032321
27
COLLECTOR ALARM (AVAILABLE WITH LCD)
Alarms
Disconnect power before removing cover. Remove the cover to set the switch settings. NOTE: the threads between the body
and the cover are very fine, so it will take multipa1le turns to remove the cover. Be very careful not to cross thread the threads.
Ensure the cover is properly re-installed and tightened prior to placing in an explosive operating environment.
Typical alarm wiring diagram is shown below:
Set High Flow Alarm
1.
Press A1 until “HFLo” is displayed, then release A1.
2. Press A2 and hold until the setpoint is displayed on the LCD (in this example, high flow
alarm is set at 80.0). Then release A2. The first digit starts blinking.