CHAPTER 3: OPERATING INSTRUCTIONS
MICROPULL
®
IV WIRE BOND PULL TESTER
3-20
990-180
2 The Reports generated by this software contain much more information than that shown above. The
Non-Destruct and Destruct Test Reports are illustrated below:
NON-DESTRUCTIVE PULL TEST
Date:
12-06-1987
14:23
Operator:
M.P.
PHFOUR
Part No:
A1234567890
Lot
No:
4-234-789
Beam Size:
100 gf
Pull Test Limit:
2.5 - 2.7 gf
No. Bonds / Device:
16
Serial
No:
1
BOND #
PULL (gf)
ERROR
NO FAILURES
Serial
Number: 1
No. of Tests:
16
No. of Good Tests:
16
No. of Bad Tests:
0
Beam Size:
100 gf
Pull Test Limit:
2.5 - 2.7 gf
No. Bonds / Device:
16
Serial
No: 2
BOND #
PULL (gf)
ERROR
1
.3 gf
NO TEST/OPERATOR
1 2.5
gf
2 2.6
gf
3 2.5
gf
4 2.6
gf
5
.5 gf
WIRE BROKE
6 2.6
gf
7 2.5
gf
8 2.6
gf
9 2.5
gf
10 2.6
gf
11 2.6
gf
12 2.5
gf
13 2.6
gf
14 2.5
gf
15 2.6
gf