Precision Pressure Measurement
The Microtector
®
Gage combines the time-
proven principles of the Hook Gage type
manometer and modern solid-state integrat-
ed circuit electronics. It provides an inexpen-
sive means of achieving accuracy and
repeatability within ± .00025 inches water col-
umn throughout its 0 to 2 inches w.c. range.
It is truly a new standard in precision measur-
ing devices.
Principles of Operation
A pressure to be measured is applied to the
manometer fluid which is displaced in each
leg of the manometer by an amount equal to
1/2 the applied pressure. A micrometer
mounted point is then lowered until it contacts
the manometer gage fluid. The instant of con-
tact is detected by completion of a low-power
A.C. circuit. Current for this circuit is supplied
by a 1.5 volt penlight cell feeding two semi-
conductor amplifiers which act as a free-run-
ning multivibrator operating at a frequency of
approximately two kilohertz. Completion of
the A.C. circuit activates a bridge rectifier
which provides the signal for indication on a
sensitive (0 to 50 microamps) D.C. microam-
meter.
2
LOW
PRESSURE
FLUID
LEVEL
OFF
ON
On indication of contact, the operator stops
lowering the point and reads the micrometer
which indicates one half the applied pressure.
By interpolating eight divisions (each being
.000125˝ w.c.) between .001 micrometer
graduations, a total accuracy of .00025 can
easily be achieved. The micrometer complies
with Federal Specification GGG-C-105A and
is traceable to a master at the NIST.
Locating and Opening
Stand the Microtector
®
Gage and case on a
firm flat level surface. Remove cover by
releasing the latches and lifting it straight up.
If it is necessary to move the gage without
case, handle only the base plate or clear
acrylic block. (
CAUTION:
Do not handle gage
by grasping meter-electronic package hous-
ing Item 7 on drawing.)
Microtector
®
Gage
D-57:D-57 7/10/09 10:39 AM Page 2
Summary of Contents for KP5522
Page 11: ...Project 22 753 Model KP5522 Page 10 Appliance Front Appliance Left...
Page 12: ...Project 22 753 Model KP5522 Page 11 Appliance Right Appliance Rear...
Page 14: ...Project 22 753 Model KP5522 Page 13 Pellet Fuel Analysis...
Page 17: ...Project 22 753 Model KP5522 Page 16 Sealed Unit...
Page 54: ......
Page 55: ......
Page 56: ......
Page 58: ...Project 22 751 Model Charnwood Skye E700 Page 16 Tunnel Schematic...
Page 102: ......
Page 103: ......
Page 104: ......
Page 129: ...2022 United States Stove Company 25 NOTES...
Page 177: ...2022 United States Stove Company 25 NOTES...
Page 272: ...24 2022 United States Stove Company NOTES...
Page 308: ......
Page 309: ......
Page 311: ......
Page 312: ......
Page 318: ......
Page 319: ......