APENDIX A - FACTORS AFFECTING FLOW MEASUREMENT ACCURACY 

 

Starflow QSD measures velocity to an accuracy of ±2% and depth to ±1% of range. This is logged to 

a resolution of 1 mm/sec and 1mm respectively.   
The purpose of the Starflow QSD system is to produce velocity data. There are many opportunities for 

errors to accrue in the process and degrade the result.  These can be reduced or eliminated by using 

the instrument properly.  Some of the more significant potential error sources follow. 
 

Alignment with Flow and Depth  

For the calibration to be valid, the transducer needs to be horizontally and vertically aligned with the 

flow.  While Starflow QSD instruments are calibrated pointing into the flow, they can be pointed 

downstream with little loss of calibration accuracy.  You may want to do this when fouling of the 

sensor face is a problem. Any angled flow in the horizontal plane will reduce the recorded velocity.   
The Starflow QSD instrument must be mounted in the water parallel to the surface for depth readings 

to measure accurately (~ +/- 10 deg), if not the depth may read inaccurately and therefore the 

recorded depth can be recording incorrectly.  
 

Instantaneous Versus “Averaged” Velocity  

When you observe Starflow QSD velocities, they will be seen to vary by 10% or more from scan to 

scan at some sites. Because Starflow QSD is very sensitive to variations in velocities, you are able to 

see the natural velocity changes in the channel.  
Although the discharge in a channel may be reasonably constant for a period of time, the velocity 

distribution is always changing. Different velocity streams wander from side to side and bed to surface 

as they progress down the channel. Turbulent swirls and eddies are carried downstream for long 

distances while they slowly decay. Hydrographers will be used to having this action partly removed by 

the mechanical inertia of a current meter and the period over which a typical measurement is timed. 

However all will have noticed that the rate of revolutions of the current meter varies during the timing 

period.  
Continual velocity logging at one location with a Starflow QSD will show these cyclic velocity 

pulsations. The characteristics will be different for different sites and will vary with discharge. Cycles 

will typically include short period fluctuations (a few seconds) overlaid on longer cyclic fluctuations (up 

to many minutes). Longer term pulsations may also be seen particularly in larger streams when in 

flood.  
When comparing Starflow QSD velocity and mechanical current meter readings, the display should be 

observed long enough to estimate the mean of the readings. The Starflow QSD will do most of this 

processing internally but if an external logger is being used to record the readings averaging could 

also be done here this will help attenuate short frequency variations. 
 

Conversion of Logged to Mean Velocity  

The measured velocity data may have to be adjusted during post processing to reflect a mean 

velocity for the channel. The factors used will be site specific and have to be determined by the 

operator.  This is done by obtaining a mean channel velocity by conventional techniques and  
 
 

Summary of Contents for 6527B

Page 1: ...EN 61000 4 4 1995 EN 61000 4 6 1996 ENV 50204 1995 Any changes or modifications to this equipment not expressly approved by the manufacturer Unidata Pty Ltd could void the user s authority to operate...

Page 2: ...al Spread 7 3 9 Received Signal Strength Indicator RSSI 7 4 0 Taking Measurements From The Instrument 8 4 1 SDI 12 Operation 8 4 2 Modbus RS485 Operation 12 5 0 Installation 16 5 1 Site Considerations...

Page 3: ...ues the Starflow QSD is able to perform in a wide range of environments It is used to measure flows and water conductivity 6537A in pipes channels and small streams It can operate in a range of water...

Page 4: ...g Clamp SS expanding turn buckle and band clips 100mm wide 100mm long 50mm expansion Use in pipes up to 600mm 24 in diameter Model 6705B Large Expanding Clamp SS expanding turn buckle and band clips 1...

Page 5: ...n 1 mm Temperature Range 0 C to 60 C Accuracy 0 5 C Resolution 0 1 C Electrical Conductivity EC 6537A ONLY Range 0 to 200 000 S cm typically 1 of measurement Resolution 1 S cm May be recorded as a 16...

Page 6: ...2 Red Power Yellow Ground Blue A Modbus Orange B Modbus Grey One Wire Use with Unidata 6515A QSD Barometric Reference Module When Unidata 6515A QSD Barometric Reference module is connected the instrum...

Page 7: ...oppler To detect water velocity an ultrasonic signal is transmitted into the water flow and echoes reflections returned from particles suspended in the water flow are received and analysed to extract...

Page 8: ...dicated depth This is corrected by subtracting the atmospheric pressure from the measured depth pressure A barometric pressure sensor is required to do this If the data logger used to read the Starflo...

Page 9: ...to determine the amount of turbulence affecting the signal quality and used to reject the measurement if the spread turbulence is too great A good flow would return a spread value of around 50 anythi...

Page 10: ...urns an acknowledge response 3 Extended Command aX is supported see below for details 4 D R1 9 Commands operate the same as a D R0 command 5 M C1 9 Commands operate the same as an M C command 6 MC CC...

Page 11: ...4 RSSI signal level no units 66 66 no units 5 Signal Spread no units 45 45 no units 10 X Axis Tilt deg 9 9 deg left tilt 11 Y Axis Tilt deg 7 7 deg forward tilt 12 Pressure Depth mm 123 123mm 13 Baro...

Page 12: ...ater Depth in mm below which Velocity will not be measured Min RSSI signal level below which Velocity will not be measured Max Signal Spread above which Velocity will not be measured 3 Velocity Ripple...

Page 13: ...4 Enable CW Velocity Measurement Disable CW Velocity Velocity 0 25 CW Velocity 0 if MIN RSSI Retain last measurement 26 Enable ve CW Velocity ve Velocity regardless of direction 27 Enable SOS Temp Cor...

Page 14: ...l succeed but indicate failure because the existing RS 485 baud rate settings are not preserved across a hard update and have been reset to the default 9600 RS485 Baud rate and Parity setting options...

Page 15: ...ho of function code sent in master message to the device Data payload 0 N bytes with response from device CRC 2 bytes with a value computed mathematically 4 2 4 Standard Message Format Read Holding Re...

Page 16: ...only ushort EC Hi Resolution Un Corrected 0009 1 Read only ushort EC Conductivity Temperature Corrected 0010 1 Read only ushort EC Hi Resolution Temperature Corrected 0011 1 Read only sshort Accelero...

Page 17: ...nused 0230 1 Read only ushort Unused 0231 1 Read Write ushort CW Velocity use half power point 0232 1 Read Write ushort CW Velocity zero above Spread Device Information and Communications Registers Re...

Page 18: ...resent in the water Generally the more material in the water the better Starflow QSD generally works well in clean natural streams but problems may be encountered in extremely clean water No excessive...

Page 19: ...hat is slipped inside the pipe and expanded to lock it in position In open channels special mounting brackets may be required The Model 6705 Expanding Band kit allows you to install a Starflow QSD ins...

Page 20: ...ow QSD mounting clamp 6527M as a guide Use a small pilot drill bit about 2mm then finish with the 5mm bit When drilling stainless steel use a slow speed and high pressure Avoid stopping as the materia...

Page 21: ...g screenshots are used to explain the functionality of the software and the various parameters that the user is able to change As with all configuration changes care must be taken to mitigate the rami...

Page 22: ...the software returns an error it may mean that the Device Driver for the interface cable did not install correctly Navigate to the computer Device Manager page and check the status of the device driv...

Page 23: ...on page for the Starflow QSD instrument This is the page where most of the settings can be directly edited The first area marked in RED is the settings associated with the MODBUS RS 485 settings QSD A...

Page 24: ...gives the instrument a unique SDI 12 address in case there is more than one SDI 12 instrument on the common SDI 12 address The setting associated with the Ultrasonic depth sensor is shown above When...

Page 25: ...a equal pressure between the inside and outside of the enclosure The settings above reflect some of the options associated the the pressure depth sensor Zero Result On Failure If this check box is tic...

Page 26: ...locity readings are returned by the QSD instrument Zero Result on RSSI fails This will return a zero velocity reading if the RSSI value Is below a certain point This value is set in the Min RSS Signal...

Page 27: ...VE any modified settings using the box Once this button has been pressed a confirmation dialog box will appear You can elect to save the changes or to cancel the operation No changes will take place u...

Page 28: ...able for standard users You may be asked by Unidata Support Staff to use this option to diagnose particular problems with the QSD Sonic COM Port This is generally set to AUTO but the drop down will li...

Page 29: ...rum is able to give a visual indication of what the Instrument can See It generates a graph that represents the flow of water passing over the instrument The more defined this graph is will determine...

Page 30: ...tion of a good Doppler signal The above example is a poor signal as the signal strength RSSI has halved and the Qutl Spread quadrupled from the previous good signal The graph peak is not well defined...

Page 31: ...cess Once the firmware update selection is made a file window will open and the firmware selection file can take place New firmware files may be sent to Starflow QSD users by the Unidata Support Team...

Page 32: ...to the above diagram for the areas where the ultrasonic eyes and conductivity sensors reside These areas must be kept clean The front face of the instrument and the area above the depth Piezo should b...

Page 33: ...anges in the channel Although the discharge in a channel may be reasonably constant for a period of time the velocity distribution is always changing Different velocity streams wander from side to sid...

Page 34: ...e speed of sound in water The factor used to scale the velocity measurement is based on the speed of sound in fresh water at 20 C see table below This velocity of sound gives a calibration factor of 0...