background image

 

6

 
Safety: 

 
Your PowerSpin C8624 centrifuge is a carefully designed and built instrument. When 
operated properly, it presents no safety hazards, however, please take note of these safety 
precautions. 

 

Lid with zero rpm lid latch (electronically operated): 

 

NEVER ATTEMPT TO OPEN THE LID UNTIL THE ROTOR HAS COME TO 
COMPLETE STOP

!

 

 

Load Balance

 

Always make sure that the load is balanced before operating the instrument

.  

     

 

It is very important that tubes with equal loads are placed opposite each other in the rotor. 
Also, when spinning a single specimen tube, another specimen tube of the same size filled 
with water must be installed directly across from the specimen tube to maintain balance. If 
the centrifuge is not properly balanced, damage could occur. If excessive noise or 
vibration is noticed during a run, stop the centrifuge and recheck the balance. 
 

Maintenance: 

 
Be sure to disconnect the power before attempting to clean. 

 

This centrifuge is designed to be relatively maintenance free. The motor is a brushless 
type and has permanently sealed bearings so oiling is not necessary. A surface cleaner 
such as Formula 409 can be used to clean the outside of the instrument. The rotor chamber 
and cover can be disinfected with a mild solution of bleach or other disinfecting liquid and 
allowed to dry. Do not drench or immerse the unit in any liquid!  

 

Calibration: 

 

It is recommended that your centrifuge’s RPM be calibrated at least every 6 months. This 
period is suggested by both the FDA and C.L.I.A. guidelines. The speed should fall within 
±100rpm of the speed display. Since the speed displayed on Unico C8624 is from a real 
time digital sensor mounted on the motor shaft, it reflects the real running speed of the 
rotor.  The timer is an electronic timer and no calibration is required. 
 

Electrical Shock Hazard

 

Do not attempt to operate this instrument near water or liquid spill. Unplug the unit before 
cleaning or repairing. Periodically check the condition of the power cord for cuts or 
damage. Do not operate with a damaged cord. 
 

Biohazard: 

 

If a tube breaks, a biological hazard may exist. Unplug the instrument. Clean using 
approved laboratory procedures. Always wear the proper protective clothing and gloves. 
Use a suitable disinfectant and dispose of the sample properly. Refer to the guidelines set 
forth by OSHA, CLIA, etc. 
 

Summary of Contents for PowerSpin C8624

Page 1: ...PowerSpin C8624 24 Place Centrifuge OPERATION MANUAL 182 Ridge Road Suite E Dayton NJ 08810 USA 800 588 9776 Fax 732 274 1151 Web www unicosci com...

Page 2: ...tal display simultaneously Electronic 60 minutes timer with beeping signal at end of run Automatic time and speed reset to the last time speed and time setting for repeated job Memorized time and spee...

Page 3: ...shion foam in the rotor chamber Do not operate the centrifuge before the foam is removed 1 Take the centrifuge from the Styrofoam and place it on a sturdy flat surface Do not turn on the centrifuge 2...

Page 4: ...conds To increase time setting press and hold up arrow otherwise down arrow button Speed Setting and Speed Setting Change Speed setting Use the up or down arrow to preset the running speed The speed r...

Page 5: ...total to date number of cycles will be displayed Note Cycles of usage is not total spinning time It represents the total times the centrifuge has spun Operation 1 Plug the centrifuge into a power sou...

Page 6: ...is not necessary A surface cleaner such as Formula 409 can be used to clean the outside of the instrument The rotor chamber and cover can be disinfected with a mild solution of bleach or other disinfe...

Page 7: ...switch in the back of unit is in the on position e Check that the power indicator light with the switch is lit A failure of the power light to illuminate could indicate a need for a new fuse when the...

Page 8: ...which is 250V 4A e Put the snap in fuse holder back 5 Intermittent motor operation or no operation Contact your representative for possible motor replacement Specifications Speed 1 000 3 400 RPM Spee...

Page 9: ...ed distributor or dealer within five 5 days after the occurrence of said defect No other warranty of any kind is made expresses or implied The warranty described above shall be the sole and exclusive...

Page 10: ...o response no response sound 1 Loose connection or connection reversed 2 Touch panel defective 1 Exercise the connector J2 on main 2 Replace touch panel 6 Display show Running but motor is not running...

Page 11: ...4 Centrifuge Parts and Nomenclature Lid Rotor Emergency lid opening left side Control panel Start Stop Open buttons Digital Timer Speed Control Safety lock On off switch on back side Power indicator M...

Reviews: