5
Stop Button
: Stop the centrifuge at anytime.
Open Button
: Open the centrifuge lid only when the centrifuge is not spinning.
Emergency Lid-opening
:
The lid is designed to prevent it from being opened accidentally while the
centrifuge is spinning. There may be situation when the lid needs to be opened
manually (For example if you lose electrical power and samples are still in the
centrifuge).
To open the lid manually, use the lid-opening tool (or a small screw driver). Insert
the tool into the emergency lid-opening hole on the up-right corner on the left side
(as showed in Installation Section). Push the tool in to release the safety lock and
open the lid.
Electronic cycle counter:
The centrifuge will count the total cycles of usage. To check the cycle usage, press
and hold Stop Button for 4 seconds, the total to date number of cycles will be
displayed.
Note
: Cycles of usage is not total spinning time. It represents the total times the
centrifuge has spun.
Operation:
1.
Plug the centrifuge into a power source outlet. Make sure the voltage is correct.
2.
Open the lid by pressing the Open button on the control panel.
3.
Place the specimen tubes into the tube shields of the horizontal rotor.
CAUTION!!
Make sure that the specimen tubes are balanced.
Always balance rotor with tubes of equal size and equal fill.
(See safety section below
)
4.
Close the lid firmly.
5.
Set the timer to the desired time by pressing the up or down arrow. To speed
up time setting, press and hold the arrow button.
6.
Press the Start button to start the spin
7.
After the centrifuge rotor has come to a complete stop (three beeping sounds),
the lid will automatically be released. Open the lid.
Note: When the centrifuge run is complete, the timer will be automatically reset to
the last timer setting and the centrifuge is ready for the next run.