Service: 

 
If you feel that your centrifuge needs factory service, please call us and speak to a 
customer service representative to determine the nature of the problem. Simple problems 
may be resolvable over the telephone. If the instrument needs to be returned to Unico, a 
R.G.A. (return goods authorization) number will be issued. Please reference the R.G.A. 
number on the paperwork packed with the instrument, and on the outside of the carton. 

 
REPAIRS OR MODIFICATIONS MADE TO THE CENTRIFUGE BY AN 
UNAUTORIZED PARTY WILL VOID THE WARRANTY. 
 
Problem Solving

 

1.

 

Centrifuge does not start: 

 

a) Check to determine if centrifuge is plugged in. 
 
b) Check that lid is properly closed. 
 

      d) Check that Power switch in the back of unit is in the “on” position. 
 

 

e) Check that the power indicator light (with the switch) is lit. A failure of the     
power light to illuminate could indicate a need for a new fuse when the unit is 
properly plugged in, switch is turned on and the lid is closed. 

 
2.

 

Excessive noise when running. If the noise is irregular or different in tone, be 
sure that there is no paper or debris in the rotor or rotor chamber. Make sure 
the load is balanced in the rotor. 
 

3.

 

Indicator light fails to light: 

 

a) Unit is not turned on or no power supply. 
 
b) Fuse is blown. 

 
4.

 

Fuse Replacement: 
 

Caution

:  

DISCONNECT THE POWER CORD BEFORE REPLACING FUSE. 

 

Follow the procedures below to replace the fuse: 
 
a)

 

Unplug the power cord. Turn the on/off switch to “off” 

 
b)

 

Remove the power cord from the power inlet on the back of the centrifuge. 

 
c)

 

Locate the fuse holder. The fuse holder is a part of the power inlet. 

 
 
 

Summary of Contents for PowerSpin C8624

Page 1: ...PowerSpin C8624 24 Place Centrifuge OPERATION MANUAL 182 Ridge Road Suite E Dayton NJ 08810 USA 800 588 9776 Fax 732 274 1151 Web www unicosci com...

Page 2: ...tal display simultaneously Electronic 60 minutes timer with beeping signal at end of run Automatic time and speed reset to the last time speed and time setting for repeated job Memorized time and spee...

Page 3: ...shion foam in the rotor chamber Do not operate the centrifuge before the foam is removed 1 Take the centrifuge from the Styrofoam and place it on a sturdy flat surface Do not turn on the centrifuge 2...

Page 4: ...conds To increase time setting press and hold up arrow otherwise down arrow button Speed Setting and Speed Setting Change Speed setting Use the up or down arrow to preset the running speed The speed r...

Page 5: ...total to date number of cycles will be displayed Note Cycles of usage is not total spinning time It represents the total times the centrifuge has spun Operation 1 Plug the centrifuge into a power sou...

Page 6: ...is not necessary A surface cleaner such as Formula 409 can be used to clean the outside of the instrument The rotor chamber and cover can be disinfected with a mild solution of bleach or other disinfe...

Page 7: ...switch in the back of unit is in the on position e Check that the power indicator light with the switch is lit A failure of the power light to illuminate could indicate a need for a new fuse when the...

Page 8: ...which is 250V 4A e Put the snap in fuse holder back 5 Intermittent motor operation or no operation Contact your representative for possible motor replacement Specifications Speed 1 000 3 400 RPM Spee...

Page 9: ...ed distributor or dealer within five 5 days after the occurrence of said defect No other warranty of any kind is made expresses or implied The warranty described above shall be the sole and exclusive...

Page 10: ...o response no response sound 1 Loose connection or connection reversed 2 Touch panel defective 1 Exercise the connector J2 on main 2 Replace touch panel 6 Display show Running but motor is not running...

Page 11: ...4 Centrifuge Parts and Nomenclature Lid Rotor Emergency lid opening left side Control panel Start Stop Open buttons Digital Timer Speed Control Safety lock On off switch on back side Power indicator M...

Reviews: