UTD2000E Series Osilloscope Programming Manual
:WAVeform:SOURce?
Functional description:
Set or query signal source of the current waveform data, if not send this command, it presents
query waveform data of the current channel.
Return format:
Query return
{ CHANnel1 |
CHANnel2 |
CHANnel3 | CHANnel4 | MATH }.
For example:
:WAVeform:SOURce CHAN1
set signal source of the current waveform data as channel 1
:WAVeform:SOURce?
query return CHANnel1
:WAVeform:POINts
Command format:
:WAVeform:POINts <points>
:WAVeform:POINts?
Functional description:
Set return waveform data, the default value is 0.
Return format:
Query return waveform point.
For example:
:WAVeform:POINts 120
set return waveform point as 120
:WAVeform:POINts?
query return 120
:WAVeform:DATA?
Command format:
:WAVeform:DATA?
Functional description:
Read waveform data of the specified data source
。
Return format:
WAVeform:POINts the specified quantity of waveform data, waveform data source is related
with :WAVeform:SOURc, data format is related with WAVeform:FORMat, returns data is
conform to binary data IEEE 488.2 # format.
For example:
The execute order to obtain the specified data source the waveform data as following:
The flow to obtain the waveform data of screen
:WAVeform:SOURce CHAN1
Set channel 1 as singal source of query waveform data
:WAVeform:MODE NORMal
Set to read the display waveform data