![Uni-Trend UTD2000E Series Manual Download Page 32](http://html1.mh-extra.com/html/uni-trend/utd2000e-series/utd2000e-series_manual_839099032.webp)
UNI-T
UTD2000E Series Osilloscope Programming Manual
Query return 5.000e-003
,
unit is s.
TRIGger Command
This command is to control trigger sweep mode and trigger standard, trigger decide
when oscilloscope start to collect data and display waveform.
Trigger Control
:TRIGger:MODE
Command format:
:TRIGger:MODE <mode>
:TRIGger:MODE?
Functional description:
Set trigger mode.
<mode> divided into EDGE
(
edge trigger
)
, PULSe
(
pulse width trigger
)
, VIDeo
(
video trigger
)
,
SLOPe
(
slope trigger
)
, ALTernat
(
alternative trigger
)
.
Return format:
Query return trigger mode
For example:
:TRIGger:MODE EDGE
edge trigger
:TRIGger:MODE?
query return EDGE
:TRIGger:SWEep
Command format:
:TRIGger:SWEep {AUTO|NORMal|SINGle}
:TRIGger:SWEep?
Functional description:
Set trigger sweep mode.
AUTO
:
If there has no trigger term, internal will produce trigger signal to force trigger.
NORMal
:
Trigger only when in trigger term.
SINGle
:
In trigger term, trigger one time and stop.
Return format:
Query return trigger sweep mode {AUTO|NORMal|SINGle}.
For example:
:TRIGger:SWEep AUTO
set channel 1 as AUTO mode
:TRIGger:SWEep?
query return AUTO