UTD2000E Series Osilloscope Programming Manual
:CURSor:YDELta?
Command format:
:CURSor:YDELta?
Functional description:
In cursor track measuring, the difference value
△
Y between cursor A point and cursor B
point.Unit is the same as the current chnnel unit.
Return format:
Query return the actual difference value
△
Y between cursor A point and cursor B point by
scientific notation method.
For example:
:CURSor:YDEL?
query return 2.02
ACQuire Command
This command is to set sampling mode of oscilloscope.
:ACQuire:TYPE
Command format:
:ACQuire:TYPE {NORMal | AVERage | PEAKdetect | HRESolution }
:ACQuire:TYPE?
Functional description:
Set
sampling mode of oscilloscope.
NORMal, AVERage, PEAKdetect and HRESolution.
Return format:
Query return {NORMal | AVERage | PEAKdetect | ENVelope | HRESolution }.
For example:
:ACQ:TYPE AVER
set sampling mode as AVERage
:ACQ:TYPE?
query return AVERage
:ACQuire:AVERages:COUNt
Command format:
:ACQuire:AVERages:COUNt <count>
:ACQuire:AVERages:COUNt?
Functional description:
Set the average sampling time during average sampling mode, <count> stepped by 2 of Nth
power, Nth take value from 2~8192, 1
≤
N
≤
30.
Return format:
Query return the current sampling time.
For example: