![Uni-Trend UTD2000E Series Manual Download Page 29](http://html1.mh-extra.com/html/uni-trend/utd2000e-series/utd2000e-series_manual_839099029.webp)
UNI-T
UTD2000E Series Osilloscope Programming Manual
:MEASure:FALLtime?
Command format:
:MEASure:FALLtime? [<source>]
Functional description:
Measuring fall time of the specified channel waveform, <source> take value as CHANnel1,
CHANnel2, CHANnel3 and MATH. Ignore means the current channel.
Return format:
Query return 5.000e-005
,
unit is s.
:MEASure:PERiod?
Command format:
:MEASure:PERiod? [<source>]
Functional description:
Measuring period of the specified channel waveform, <source> take value as CHANnel1,
CHANnel2, CHANnel3 and MATH. Ignore means the current channel.
Return format:
Query return 5.000e-003
,
unit is s.
:MEASure:PWIDth?
Command format:
:MEASure:PWIDth? [<source>]
Functional description:
Measuring period of the specified channel waveform, <source> take value as CHANnel1,
CHANnel2, CHANnel3 and MATH. Ignore means the current channel.
Return format:
Query return 5.000e-003
,
unit is s.
:MEASure:NWIDth?
Command format:
:MEASure:NWIDth? [<source>]
Functional description:
Measuring negative pulse width of the specified channel waveform, <source> take value as
CHANnel1, CHANnel2, CHANnel3 and MATH. Ignore means the current channel.
Return format:
Query return 5.000e-003
,
unit is s.