![Uni-Trend UTD2000E Series Manual Download Page 13](http://html1.mh-extra.com/html/uni-trend/utd2000e-series/utd2000e-series_manual_839099013.webp)
UNI-T
UTD2000E Series Osilloscope Programming Manual
LED light staus
:
0 presents light off, 1 presents light on.
For example:
:KEY:AUTO
set oscilloscope control parameter automatically
:KEY:AUTO:LOCK ON/OFF
lock/unlock button
:KEY:AUTO:LOCK?
query return this key lock status, 1 presents locked
:KEY:AUTO:LED?
query return LED status
,
0 presents light off
CHANnel Command
This command is to set channel independently. <n> take value as 1/2/3/4/5/6/7/8/9, it
presents {CH1/CH2/ CH3/ CH4/ MATH/ REF-A/ REF-B/ REF-C/ REF-D}.
Notes: this command in instruction set is not only valid for physical channel, but also
valid for MATH and reference channel parameter.
:CHANnel<n>:BWLimit
Command format:
:CHANnel<n>:BWLimit {{1|ON}|{0|OFF}}
:CHANnel<n>:BWLimit?
Functional description:
Set bandwidth limit function is ON (turn on bandwidth to 20MHz
,
to reduce display noise) or
OFF (turn off bandwidth to full display
)
Return format:
Query return 1 or 0
,
it presents ON or OFF.
For example:
:CHAN1:BWL ON
turn on channel 1 bandwidth limit
:CHAN1:BWL?
query return 1
,
it presents channel 1 bandwidth limit is turn on
:CHANnel<n>:COUPling
Command format:
:CHANnel<n>:COUPling {DC|AC|GND}
:CHANnel<n>:COUPling?
Functional description:
Set channel coupling mode, DC presents AC and DC component of input signal; AC presents DC
component of blocking input signal; GND (grounding) presents break input signal.
Return format:
Query AC, DC or GND.
For example:
:CHAN1:COUP DC
set channel 1 coupling mode as DC
:CHAN1:COUP?
query return DC