Definitions :
Normal sampling mode :
The digital storage Oscilloscope acquires signal sample at equal and regular intervals to
reconstruct the waveform.
Peak detect mode :
In this acquisition mode, the digital storage Oscilloscope identifies the biggest and smallest
values of the input signals at each sampling interval and uses these values to display the waveform. In effect, it can
acquire and display narrow pulse which would otherwise be omitted in the sampling mode. Noise seems to be more
significant in this mode.
Average mode :
The digital storage Oscilloscope acquires several waveforms and takes the average value to display
the final waveform. You can use this mode to reduce random noise.