UTD1000 User Manual
Note : When using the cursor to measure voltage,
follow step 2 only and set the cursor type to voltage.
Illustration 6 : Using the X-Y function
To check the phase difference between signals of the
two channels.
Example : To measure the phase variation when
a signal passes through the circuit, connect your
Oscilloscope to the circuit and monitor the input and
output signals. To check the circuit’s input and output in
the X-Y coordinate mode, follow the steps below :
1.Set the attenuation factor of the channel probe to 10X.
Set the switch of the probe to 10X.
2.Connect the Channel A probe to the input terminal of
the network. Connect the Channel B probe to the output
terminal of the network.
3.If the channel is not on, press the [A] and [B] menu
button then press [F1] to enable the two channels.
4.Press [AUTO].
5.Adjust the vertical amplitude range to make the
displayed amplitude of the two channels about equal.
6.Press the [SCOPE] menu key then [F3] to recall the
display control menu.
7.Press the jog dial to select X-Y. The Oscilloscope will
Figure 3-6 Measuring pulse width with the cursors