Note: 

     Do not measure AC power supply over 750Vac or DC power supply over 

     1000Vdc. It is possible to display higher voltage (10%), but it may damage the 

      tester.

      Avoid electric shock when working with high voltage.

      After completing all measurement operations, please disconnect the test 

      lead with the measured circuit, and remove the test lead from the input 

      terminal.

Black

Red

Capacitor

9.3.1.5 Comparison Measurement

Press “COMP” button under insulation resistance testing condition without high 

voltage output. When the LCD shows “COMP”, it indicates the tester enters 

comparison measurement mode. The default comparison resistance is 10M

Ω

the initial interface flashes at the unit of 10M

Ω

 at a frequency of 1HZ, to indicate 

the tester is in comparison resistance setting state. Press       and      to select 

the digit and unit of comparison resistance to be changed, press  

  

and      to 

adjust comparison resistance and unit, then press “ENTER” to save the parameter 

setting or press “ESC” to cancel the parameter setting. After that, hold down 

“TEST” for 2 seconds, if insulation resistance is less than comparison resistance, 

the symbol “FAIL” will appear on the LCD, otherwise “PASS” appears.

To return to the continuous measurement interface, please press “COMP” in 

“COMP” mode or press “ESC” twice consecutively. 

9.3.2 Voltage Measurement

1) Connect red test lead with “LINE” input terminal, and black with “EARTH”.

2) Connect red and black alligator clips with measured circuit. For DC voltage 

    measurement, if the voltage of red test lead is negative, the negative symbol 

    “-” appears on the LCD. 

Fig. 9.3.2.1 Wire connection of voltage measurement

Black

Red

As a part of insulation measurement, the tester has circuit capacitance measurement 
function. When CAP/R button is pressed without high voltage output, the tester 
enters capacitance measurement function by default, with output voltage at 250V 
range. Under capacitance measurement function, there are only three voltage 
ranges including 250V, 500V and 1000V, which can be switched by pressing ▲ 
and ▼. For capacitance measurement, the tester calculates the capacitance by 
measuring the charge and voltage of measured circuit under charge.

9.3.3 Capacitance Measurement (UT512E)

Fig. 9.3.3.1 Wire connection of capacitance measurement

UT512D/E User Manual

UT512D/E User Manual

Summary of Contents for UT512D

Page 1: ...UT512D UT512E 2 5kV Insulation Resistance Tester User Manual P N 110401111817X...

Page 2: ...Trend guarantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misu...

Page 3: ...ainly used to measure insulation resistance of cable motor power generator transformer mutual inductor high voltage switch lightning arrester and more it is an ideal measurement meter for electric pow...

Page 4: ...T512E 8 PCS LR14 alkaline batteries UT512D Error limit a of reading b digits one year warranty Environment temperature 23 5 Environment humidity 45 75 RH In the table below The humidity must be less t...

Page 5: ...ications UT512E Function Measurement range Accuracy Remark Low resistance measurement Open circuit voltage 5V Note If the measured resistance is 20 the buzzer alarms if 660 660 is displayed The short...

Page 6: ...bidden to use the tester without battery cover set in place otherwise it may pose a risk of electric shock Keep fingers behind the finger guard when perform measurement do not touch exposed wire conne...

Page 7: ...plug red lead GUARD Grounding protection outlet single plug green lead EARTH High resistance measurement outlet single plug black lead LCD segmented display screen Functional buttons USB port Carryin...

Page 8: ...testing Bluetooth Insulation resistance measurement mode AC DC voltage measurement mode Power on off Increase select high range or previous group of data Decrease adjust time and resistance cyclically...

Page 9: ...ress this button to switch to insulation resistance testing function V UT512E AC DC voltage testing button Without high voltage output short press V button to switch to AC DC voltage testing mode The...

Page 10: ...measurement Data deletion button In READ state short press CLR and then press ENTER to delete current data press ESC to exit data deletion function In READ state the symbol CLR and the storage capaci...

Page 11: ...rged if low battery occurs please replace with battery in same type 9 2 Battery Charging UT512E When charging with battery charger in power on state the battery power indicator and the charging symbol...

Page 12: ...hange Press TEST to perform test the LCD shows basic elements including battery power high voltage warning symbol leakage current real time output voltage measured insulation resistance testing value...

Page 13: ...tion resistance in Time1 9 3 1 4 Dielectric Absorption Ratio Measurement Dielectric Absorption Ratio DAR refers to the value of insulation resistance in 1 minute to that in 15 seconds It takes 1 minut...

Page 14: ...s if insulation resistance is less than comparison resistance the symbol FAIL will appear on the LCD otherwise PASS appears To return to the continuous measurement interface please press COMP in COMP...

Page 15: ...ltage input the danger symbol flashes in such case do not press the test button to measure resistance as shown in figure 9 3 4 2 Under resistance measurement function connect the black grounding lead...

Page 16: ...rminal with the exterior conductor of inner insulation layer and the unused cable Surface leakage current is conducted to the safety terminal so as to eliminate the surface leakage current at the meas...

Page 17: ...Red XI Maintenance Clean the casing 1 Wipe the surface using soft cloth or sponge dipped with water 2 Do not immerse the tester in water to avoid tester damage 3 Do not store the tester if it is wet 4...

Reviews: