NT228B GB2
28 / 98
6.1
Mode « Simple »
This menu is dedicated to flow measurements on water, at ambient temperature,
with a minimum of settings.
To activate this mode, select it in the configuration menu. You keep the possibility
to add other optional functions (I/O, data logger …).:
With Simple Option Menu, some settings, which are adjustable with Normal or
Advanced option, are here forced at following values:
Product = Water with Co=1472 m/s and Delta C=+/- 400 m/s
K Hydro = Auto / Viscosity = 1.0 CSt / Pipe Roughness = 0.1 mm
Flow Cut-Off = 0 / Auto Zero = No
Memory = 10 s / Filter = 10 s
Graph Period = 10 s
Supposing your instrument already in that mode, after switching ON, the UF 801-P
displays menu title, the instrument configuration and serial numbers (see 3.3) and
then the chord / probes configuration (see 6.4).
UF 801-P
1 Pipe
Date : 01/08/2007
Serial/N: 01/99/2007
Hardware: MK5-E
Software: 01-12-01-A
Otherwise, a long press on
F
switches the device to measurement menu.