NT228B GB2
69 / 98
------ TIMER -------
! Data-Logger !
------Recording-----
Log. Step= ****
Duration = **h**m**s
----Powering On-----
Start = **h**m
Period = **h**m
---Total Duration---
99999 d *99 h 99 mn
11:10
Very important
When your selected mode is set, you must valid it by a long press on F to returns
to the Measurement menu.
Check everything is OK concerning your installation and measurement conditions.
Then, switch OFF the UF 801-P/PB.
The UF 801-P will self-switch ON at set time and then periodically (if requested) as
per your settings.
Complementary information
If you have set the Differed mode, the Timer is automatically reset OFF after the
operation.
If you have set the ON mode: the Timer is automatically reset after the set whole
time is over.
The Infinite mode must be reset manually. You can do this when it is ON.
Otherwise, proceed as below:
In any case, if the timer is started and the UF 801-P is waiting for its automatic
start-up, pressing the ON button will restart the UF 801-P and give you the
possibility to cancel or modify all settings that you will need to valid as before.