MD40
PRODUCT INSTRUCTION MANUAL
MD40
Page 16 of 56
Author: N.Carapiet | Revision: 1.1
Refitting the Drive Thimble
1.
Replace Motor Flange Clamp and tighten 6 x M4 Motor Flange Clamp Screws.
2.
Push Drive Thimble over the Flange Assembly (take care when re-coupling the magnets
–
the magnetic coupling may cause a sudden movement)
3.
Replace Bearing Spacer on Thimble.
4.
Replace Thimble Cover and tighten 3 x M3 Grub Screws
3 x M3 Grub Screws
Thimble Cover
Thimble
Motor Flange Clamp
M4 Screws x 6
Motor Flange Clamp
Bearing Spacer