Instructions for use (rev 011221)
e.ON sinter fast +
UGIN DENTAIRE
23
8.
MAINTENANCE
Before to carry out the maintenance work and/or cleaning, e.ON sinter fast + has to:
-
disconnected from the main power supply
(switch OFF and unplug to mains socket).
-
at room temperature: wait the cooling after use
(risk of burns and/or damage).
Observe the following recommendations to ensure the longevity of the device:
-
Do not use abrasive, solvents or flammables products for cleaning (outside and inside).
-
Clean the machine outside with a soft, dry cloth.
-
Remove any traces on the
screen
with a soft cloth and a small amount of special glass cleaner. Cleaning
wipes for optic products can also be used. For avoid screen damage, do not use a sharp object to operate
the touchscreen.
-
Remove any pearl of small pieces stuck between the display and the furnace casing.
-
In case of dust into the
chamber
: remove it with a vacuum cleaner
–
do not used compressed air.
-
Visual inspection of the support and the tray
before use.
If necessary, clean with a vacuum cleaner to remove any particles (dust) - do not use compressed air.
If the support is damaged or cracked: do not use and replace-it.
-
Regularly check the
electrical contacts
of the heating elements and tighten (every two weeks).
-
Lubricate the carriage carrying the tray regularly.
NOTE: do not use compressed air to cleaning: risk of damage to the accessories and the suspension of particles
of the refractory material.