36
7.
Place the rod in the middle of the transducer are (as above)
8.
Select a volume level on the Warble Tone Generator (Low or High).
9.
Note the level (L or H) in the TG box of the Sensitivity Validation chart.
10.
Turn the Ultraprobe 15,000 on
its’
side so that it will rest flat on the test table
with the handle facing the user and the Scanning Module facing the Tone
Generator.
11.
Slide the Ultraprobe gently so that the front faceplate touches the Rod and that the
rod is touching the face plate while touching the side of the Scanning Module. Align
the Scanning Module so that the center of the module is facing the center of the
Tone Generator Transducer
(on the right).
12.
Adjust the sensitivity until the
intensity bar graph is at mid-line
and displays the decibel level.
13.
Note and record the decibel
reading in the dB box of the
Sensitivity Validation chart.
B.
For the Contact (Stethoscope) Module, insert the Module into the Front End
of the Instrument
.
1.
Select 40 kHz as the test frequency and note “40” in the
Frequency box for the
Contact Module in the Sensitivity Validation Chart.
2.
Plug in the Headphones and adjust the ear pieces so that they are opened up and
place them on the test table.
3.
Place the Warble Tone Generator flat facing up with the recharge jack facing the
user at 90o.
4.
Select a volume level on the Warble Tone Generator (High or Low).
WTG on Side Recharge Jack on Left,
Volume Control on right
ROD