15
inspection. If no or very little ultrasound is present of if the
sensitivity value is too high for the test area, the dB will not
show on the screen. 3 Dashed (---) lines will show. Adjust the
“S” (Sensitivity value) by touching the Sensitivity box and then
use the Up/DOWN arrows to move the S value up or down as
needed.
d.
To freeze a reading for saving or observation, release the
trigger.
1.
Home:
When the Ultraprobe is turned on, the Home screen will be displayed. There are 4
icons shown: Main screen, Setup, dB and Temperature Screen, and Route. Should the Valve
or Steam application be selected, the Valve icon (ABCD) will also be shown. Select one of
the icons to enter and use the full features of the Ultraprobe 15,000.
NOTE
: To change from one operating screen (such as Main, dB/Temp or Valve) to another,
you must return to the Home screen.
2.
Main (dB):
This screen will display the Record number, Decibel, Frequency, Sensitivity
Level and an Intensity Level box. The Intensity Level box also acts as the Sensitivity
Control. This will be explained below, there are two control arrows on the right to be used
to adjust or change the sensitivity, frequency, record number and emissivity.
a.
To change the Frequency: touch the kHz (frequency), when a highlight appears
around the selected area then use the UP/DOWN arrows to adjust.
b.
To change the Sensitivity: touch the sensitivity value, when a highlight appears
around the selected area then use the UP/DOWN arrows to change. You will see the
change in the sensitivity value in the upper left of the screen. An alternative to using
the arrows is to use the intensity level box, which uses a bar graph to indicate
intensity. Tap the box in either the high (right) or low (left) area of the box until the
Sensitivity value (S=) changes as needed).
c.
To change to different Record Number: touch the Record Number, when a highlight
appears around the selected area then use the then use the UP/DOWN arrows to the
desired location.
d.
To save the data, touch the bottom of the screen and locate the save icon, touch save,
when prompted, touch yes to save.
Viewing changes in Decibel Levels
:
To view decibel levels the Ultraprobe must be in the active scan mode
. Pull the
trigger in to activate the active scan mode. To freeze the data for storage, while
pointing or touching the test point in the active mode, release the trigger. The data
will be frozen on the screen until it is either stored or the Trigger is pulled in.