23
ALARM ENABLE/DISABLE
While in the Operation mode, this function selection will display Alarm Disabled if no
alarm level has been set. If an alarm level has been set, it will show the set level in dB.
To enable the alarm, see ALARM ENABLE/DISABLE in the SETUP MODE below.
WARNING:
If the alarm value is changed, the new dB alarm value is stored, and will
overwrite the previous value.
The alarm dB level stored in the current record can be viewed in the main display. It can
be changed. To do so:
Click down to the bottom of the Display Panel until
“PRESS ENTER FOR” is blinking
.
Spin to Alarm (the dB value will be shown).
Click to Enter.
Spin to the desired dB alarm level.
Press the Yellow ENTER Button to exit.
GENERIC INFO
To view Generic Info, click the Sensitivity Control Dial to the bottom of the Display
Panel until “Press ENTER For” is blinking. Spin to Generic Info and press Enter.
To enter a test condition: Spin to Test the condition default is OK and will blink slowly.
If it is necessary to change:
Click with the Sensitivity Control Dial and the condition (OK) will blink rapidly.
Spin to select a condition indication. The choices are; OK, CHK and it is also possible
to customize the results codes for other conditions. These codes can be changed in
the Ultratrend software and uploaded to the Ultraprobe.