KANGAROO JOEY
™
32
Enteral Feeding Pump
section X — specifications and symbols
specifications
Ul listing
Medical equipment
KANGAROO JOEY™ Enteral Feeding Pump
(1) Classified with respect to electrical shock, fire, and mechanical hazards
in accordance with UL60601-1.
(2) Classified with respect to electrical shock, fire, mechanical and other
specified hazards in accordance with CAN/CSA C22.2 No. 601.1
Type Infusion Device
Volumetric
Pumping Mechanism
Rotary Peristaltic
Pump sets
All sets operating on the KANGAROO ePump Enteral Feeding Pump will also operate on the KANGAROO JOEY Enteral Feeding Pump.
feeding formula Delivery Rate
1-400 mL/hr in 1 mL increments
feeding formula VTbD
1-3000 mL in 1 mL increments
bolus Volume
1-3000 mL in 1 mL increments
number of boluses
1-99
bolus Interval
1-24 hours in 1-hour increments
flushing solution Dose Range
10-500 mL in 1 mL increments
flushing solution Interval Range
1-24 hr in 1 hr increments
accuracy
7% or 0.5 mL/hr, whichever is larger, with top of the fluid column at 15.24 cm (6”) ± 0.76 cm (0.3”) above the top of the closed
pump door, at a room temperature of 22°C ± 2°C (72°F ± 3°F), using a new ePump feeding set for no longer than the recommended
24 hours of maximum usage. Confidence intervals for accuracy are based upon those included in the ANSI/AAMI ID26-1992 American
National Standard for Infusion Devices. Reported accuracy is based upon the time-based accuracy calculation in this standard, as
applicable for enteral feeding. Since feeding accuracy of the KANGAROO JOEY Enteral Feeding Pump is based upon formula and
not water, a 4% accuracy shift should be applied when testing the KANGAROO JOEY Enteral Feeding Pump using water. (Water will
appear to overdeliver by 4%.)