![Tyco Kendall Kangaroo Joey Operating And Service Manual Download Page 23](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-joey/kendall-kangaroo-joey_operating-and-service-manual_1186011023.webp)
KANGAROO JOEY
™
19
Enteral Feeding Pump
Hold
Pump operation can be paused by pressing
4
HolD
while it is
RUNNING
, such as in Figure 15A. The status line will indicate
HOLDING and the yellow LED indicator will be lit up to give a quick visual indication of an informational status in a darkened room.
Note: the pump’s alarm will sound if it has been left
HOLDING
for 10 minutes without further input.
There are five options available during
HOLDING
. See Figures 18A and 18B
.
figure 18a.
HOLDING
mode
options, continuous feeding mode.
figure 18b.
HOLDING
mode
options, intermittent feeding mode.
Clear Volume fed
Press
4
CleaR VolUMe
to re-zero the output feed totals “
ml feD
,” “
bolUses feD
,” and “
ml flUsH
.”
adjust settings
Press
4
aDJUsT seTTInGs
to return to one of the opening menus shown in Figures 7A-7D, thus allowing any setting to be modified.
If the pump set has been removed after going into
HolD
, the opening menu will show
LOAD A SET
, as seen in Figures 5A-5B
.
Resume In __ Minutes
(Not Available in EZ Pump Mode)
Press
4
“
ResUMe In __ MIn
” once to immediately place the KANGAROO JOEY Enteral Feeding Pump into a 30-minute resume
running mode. Pressing the
4
ResUMe In __ MIn
button again will cyle the “resume in” time back to zero. No other input is needed
to start the countdown, as the time immediately counts down after the 30-minute value is reached.
Run
Press
4
RUn
to immediately restart pump operation. This option will not be available if the pump set is removed during
HolD
.
History
(Not Available in EZ Pump Mode)
Press
4
HIsToRY
to access the
HISTORY
screen. Use this screen to review feeding and flushing history. See subsection “More
Options” below for further information.