KANGAROO JOEY
™
22
Enteral Feeding Pump
CeRT InTV
indicates the number of times the pump has been powered on since the last re-certification. See Section VI for details of
re-certifying the pump.
bioTech options
figure 22. bioTech option screen,
with lock setting, eZ PUMP MoDe,
and Communications features
activated. (full Communications
feature to be added at later date.)
loCK seTTInGs — Program locked
To “lock” the settings against unauthorized changes, press
4
loCK seTTInGs
to get the check mark, as shown in Figure 22. The
settings can only be changed again by pressing
4
loCK seTTInGs
to remove the check mark. If the settings are locked and a change
is attempted, the screen will indicate
PRoGRaM loCKeD
, and will not allow access to the parameters, Figure 16.
The pump can be
PRIMeD
or set to
RUn
,
HolD
, etc. when settings are locked.
eZ PUMP MoDe
Press
4
eZ PUMP MoDe
to activate the
eZ PUMP MoDe
feeding option. A check mark will appear next to the
4
eZ PUMP MoDe
button to indicate that the feature is active. Pressing the
4
eZ PUMP MoDe
button again will remove the check mark indicating
that the feature is off. If the
eZ PUMP MoDe
feature is turned on, the
eZ PUMP MoDe
icon will appear in the lower right corner of
the LCD next to the battery icon (see Figure 9). See page 13 for
eZ PUMP MoDe
details.