![Tyco Kendall Kangaroo Joey Operating And Service Manual Download Page 32](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-joey/kendall-kangaroo-joey_operating-and-service-manual_1186011032.webp)
KANGAROO JOEY
™
28
Enteral Feeding Pump
Turn the power off and back on to try to clear the error. If the error cannot be resolved in a timely manner, press
PoWeR
to stop
operation of the pump and put a different pump into service.
Hold error
(Adjustable Audible Volume)
The
HolD eRRoR
screen, Figure 26, will appear if the pump has been inactive, without input, for more than 10 minutes. See
subsection “Hold” for description of
HOLDING
mode.
Press
4
ConTInUe
to return to the previous screen. Where settings can be adjusted, the pump can be set to run immediately, or the
pump can be set to run in a specified number of minutes. See Figures 18A-18B to see the options on the
HOLDING
menu screen.
If the error cannot be resolved, press
PoWeR
to stop operation of the pump and put a different pump into service.
figure 26. Hold error screen.
Rotor error
(Adjustable Audible Volume)
The
RoToR eRRoR
screen appears during
RUNNING
or
PRIMING
, when the pump has detected an unusual operating condition for
the rotor. See Figure 27. The
RoToR eRRoR
generally results from a problem with the pump set tubing around the rotor.
Check that the pump set is not damaged, and re-load the tubing around the rotor as described in subsection “Loading Pump Sets”.
Press
4
ConTInUe
to restart.
If the error cannot be resolved, press
PoWeR
to stop operation of the pump and put a different pump into service.
figure 27. Rotor error screen.
Re-load the set tubing and press ConTInUe.
feed error
(Adjustable Audible Volume)
The
feeD eRRoR
screen appears when the enteral formula is no longer being delivered because the bag is empty or because of
a clog between the pump and the bag. Check the feed line to find the occlusion causing the blockage. If the occlusion cannot be
cleared, load a new pump set, prime it, and press
4
ConTInUe
to restart the feeding, Figure 28.
If the error still cannot be resolved, press
PoWeR
to stop operation of the pump and put a different pump into service.