6.4
Program 4: Liquid B -- Waste
For liquids in
sealed bottles
when the manufacturer recommends
autoclaving at temperatures of 121
C (250
F).
Nominal parameters default settings
Sterilization temperature: 121
C (250
F)
Sterilization time : 30 minutes
Operations Sequence
Insert water to chamber and heating by actuation of electrical heaters
until reaching the high preset air pulse.
Heating by actuation of electrical heaters until the sterilization
temperature is reached.
Sterilization temperature is maintained constant for the preset
sterilization time.
EL models
Slow exhaust; steam is slowly exhausted from the chamber, until it
reaches the required end temperature (approx. 85ºC) and the
pressure equals the atmospheric pressure.