Page 12 of 75 Pages
Attention:
A switch or circuit-breaker must be included in the building
installation. This switch or circuit-breaker shall be in close proximity
to the equipment, within easy reach of the operator; and marked as the
disconnecting device for the equipment.
The electrical net must be protected with a current leakage safety
relay.
The electrical network must comply with local rules or regulations.
Verify that there is an easy access to the main power switch, to the
water cut-off valve and to the current leakage safety relay.
Make sure while placing the autoclave, to leave space around the
machine, to give the technician access to service the machine.
All water connections to autoclave must be performed through
"BACK FLOW PREVENTION SYSTEM" only, as per IEC 61770.
3.10 Symbol Description
Caution! Consult accompanying documents
Caution! Hot surface.
Caution! Hot steam.
Protective earth (Ground
)
On-Off