www.shanghaidelan.com
40
9.18
Replacing the circuit breaker
Caution!
Before starting, disconnect the instrument from the power source.
1. Remove the autoclave cover (see para. 9.2
“
Dismantling the Outer
Covers of the Autoclave
”
).
2. Disconnect the wires from the circuit breaker.
3. Remove the four screws connecting the circuit breaker to the panel
(1).
4. Replace the circuit breaker with a new one.
5. Reconnect the electrical wires.
6. Reassemble the cover.
7. Turn on the autoclave and verify it operates correctly.
8. Move the circuit breaker’s lever to the “tripped” position and verify
that the autoclave turns off.
Make sure that the correct circuit breaker is installed as marked in
para.
9.19 (Fuses and Circuit Breaker Data)
!
1
Summary of Contents for 1730EK
Page 2: ...w w w s h a n g h a i d e l a n c o m...
Page 29: ...w w w s h a n g h a i d e l a n c o m 27 Electronic box 1730 Electronic box 2340 2540 7...
Page 30: ...w w w s h a n g h a i d e l a n c o m 28 Electronic box 3140 3850 3870...
Page 78: ...w w w s h a n g h a i d e l a n c o m 76 FRONT VIEW Model 2340 2540 Model 1730...
Page 80: ...w w w s h a n g h a i d e l a n c o m 78 VESSEL ASSEMBLY...
Page 82: ...w w w s h a n g h a i d e l a n c o m 80 OUTER CABINET ASSEMBLY...
Page 103: ...w w w s h a n g h a i d e l a n c o m 101 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Page 104: ...w w w s h a n g h a i d e l a n c o m 102 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Page 105: ...w w w s h a n g h a i d e l a n c o m 103 PIPING DIAGRAM FOR MODELS EA AND EKA...
Page 106: ...w w w s h a n g h a i d e l a n c o m 104 ELECTRICAL WIRING DIAGRAM MODELS 1730 120 230V...
Page 107: ...w w w s h a n g h a i d e l a n c o m 105 ELECTRICAL WIRING DIAGRAM MODELS 2340 2540E 120 230V...
Page 109: ...w w w s h a n g h a i d e l a n c o m 107 ELECTRICAL WIRING DIAGRAM MODELS 3140E 230V...
Page 110: ...w w w s h a n g h a i d e l a n c o m 108 ELECTRICAL WIRING DIAGRAM MODELS 3870 3850E 120 230V...