www.shanghaidelan.com
56
Corrections
1.1
See symptoms 22, 2
4 and
25 in
para. 11
.3.2.
2.1
Close the doo
r more tightly
or set the do
or
switch more steadily.
3.1
If amount of
water is not correct, clean the
electrode and set the water
level
3.2
See symptom 29 in para. 11.3.2.
3.3
The thermostat must be set
at a higher
switching poi
nt and the tim
e after the
electrode touches the water be prolo
nge
d.
3.4
Clean or replace w
ater inle
t valve and
exhaust valve.
4.1.1
Clean or replace el
ectrode.
4.1.2
Level
autoclave.
4.2
See symptom 18 in para. 11.3.2.
Pos
sib
le ca
use ch
eck
-u
p and tes
ts
1.1
The problem is the incorrect reading of the
temperature s
ensor.
2.1
The pressure
in the chamber rises during the
heating stage and is pressing on the door. If
the door swit
ch is lightly
pressed it could be
turned off.
3.1
The process f
ails if the elec
trode is above the
water surfa
ce
and the safety thermostat
switches off.
Check if the amount of water in the chamber
is over 350cc.
3.2
If the water quantity re
mains insufficient, the
problem is a wrong electrode signal.
3.3
Process fail
s despite correct quantity of
water.
3.4
Presence of d
irt inside the
water inlet va
lve
may cause slowdown of t
he flow.
4.1
The heating ti
me is very long and t
he heating
rate is normal
.
4.1.1
Excess quantity of water in the
chamber due to dirty electrode,
loose electrode connections or
faulty
electrode.
4.1.2
Excess quan
tity o
f water in the
chamber due to un-levelled
autoclave.
4.2
Heating elements damaged
.
Heating Stage
Symp
to
m
1.
Pr
ocess fai
ls whil
e
initiating the
heati
ng stage.
mess
ag
e H
IGH TEMP is
displayed.
2.
Message DOOR UNLOCK
is displayed during the
heating stage, although the
door is closed.
3.
LOW WAT
ER mes
sage is
displayed during the
heating stage
and process
fa
ils
.
4.
Mess
ag
e L
O
W HEAT is
displayed during the
heating stage.
Summary of Contents for 1730EK
Page 2: ...w w w s h a n g h a i d e l a n c o m...
Page 29: ...w w w s h a n g h a i d e l a n c o m 27 Electronic box 1730 Electronic box 2340 2540 7...
Page 30: ...w w w s h a n g h a i d e l a n c o m 28 Electronic box 3140 3850 3870...
Page 78: ...w w w s h a n g h a i d e l a n c o m 76 FRONT VIEW Model 2340 2540 Model 1730...
Page 80: ...w w w s h a n g h a i d e l a n c o m 78 VESSEL ASSEMBLY...
Page 82: ...w w w s h a n g h a i d e l a n c o m 80 OUTER CABINET ASSEMBLY...
Page 103: ...w w w s h a n g h a i d e l a n c o m 101 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Page 104: ...w w w s h a n g h a i d e l a n c o m 102 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Page 105: ...w w w s h a n g h a i d e l a n c o m 103 PIPING DIAGRAM FOR MODELS EA AND EKA...
Page 106: ...w w w s h a n g h a i d e l a n c o m 104 ELECTRICAL WIRING DIAGRAM MODELS 1730 120 230V...
Page 107: ...w w w s h a n g h a i d e l a n c o m 105 ELECTRICAL WIRING DIAGRAM MODELS 2340 2540E 120 230V...
Page 109: ...w w w s h a n g h a i d e l a n c o m 107 ELECTRICAL WIRING DIAGRAM MODELS 3140E 230V...
Page 110: ...w w w s h a n g h a i d e l a n c o m 108 ELECTRICAL WIRING DIAGRAM MODELS 3870 3850E 120 230V...