9.11
Cleaning and Replacing the Water Level Electrodes
The water level electrode is located at the rear bottom area of the
chamber.
9.11.1
Replacing
Caution
Before starting, disconnect the instrument from the power
source and ensure that there is no pressure in the autoclave.
Allow the autoclave to cool before removing outer covers.
1.
Remove the rear cover.
2.
Disconnect the wire from the electrode connection.
3.
Open the nut that tightens the electrode.
4.
Insert a new electrode and tighten the fixing nut to avoid
leakage.
5.
Reconnect the wire to the electrode.
6.
Test the unit.
9.11.2
Cleaning
1.
Pull out the trays and tray holder.
2.
Clean the electrode tip.
3.
Test the unit.
Water level electrode assembly
Cat No. CMT196-0004
Summary of Contents for 1730EK
Page 2: ...w w w s h a n g h a i d e l a n c o m...
Page 29: ...w w w s h a n g h a i d e l a n c o m 27 Electronic box 1730 Electronic box 2340 2540 7...
Page 30: ...w w w s h a n g h a i d e l a n c o m 28 Electronic box 3140 3850 3870...
Page 78: ...w w w s h a n g h a i d e l a n c o m 76 FRONT VIEW Model 2340 2540 Model 1730...
Page 80: ...w w w s h a n g h a i d e l a n c o m 78 VESSEL ASSEMBLY...
Page 82: ...w w w s h a n g h a i d e l a n c o m 80 OUTER CABINET ASSEMBLY...
Page 103: ...w w w s h a n g h a i d e l a n c o m 101 PIPING DIAGRAM FOR MODELS 1730 E AND EK...
Page 104: ...w w w s h a n g h a i d e l a n c o m 102 PIPING DIAGRAM FOR MODELS E AND EK EXCEPT 1730...
Page 105: ...w w w s h a n g h a i d e l a n c o m 103 PIPING DIAGRAM FOR MODELS EA AND EKA...
Page 106: ...w w w s h a n g h a i d e l a n c o m 104 ELECTRICAL WIRING DIAGRAM MODELS 1730 120 230V...
Page 107: ...w w w s h a n g h a i d e l a n c o m 105 ELECTRICAL WIRING DIAGRAM MODELS 2340 2540E 120 230V...
Page 109: ...w w w s h a n g h a i d e l a n c o m 107 ELECTRICAL WIRING DIAGRAM MODELS 3140E 230V...
Page 110: ...w w w s h a n g h a i d e l a n c o m 108 ELECTRICAL WIRING DIAGRAM MODELS 3870 3850E 120 230V...