Cancel calibration
– causes the loss of all currently created data points. If
this button is
clicked
you will have to confirm the operation that has been
chosen.
3.7.5 View Graph
This section allows users to choose the type of equation to fit through their
data points, to force fit the equation, view the equation chosen, print the
graph, accept the equation, or go back to add data points.
A minimum of two points is required to access the View Graph screen.
Type of Equation
– defines the number of terms used in creating the formula.
You can choose from 1
st
, 2
nd
, or 3
rd
Order equations while viewing the
graphed results and the curve will shift accordingly.
Force thru First
– allows users to force the calibration curve through the first
calibration point on the curve (point closest to zero).
View Equation
– displays the equation Order and calibration coefficients for
the calibration curve being viewed.
Add Data Points
– allows you to go back to the previous screen and continue
to store more calibration points or cancel the calibration being performed.
Accept Equation
– c
lick
the Accept Equation button to accept the calibration
and store calibration data for that parameter group in the DataBank’s
memory. Upon clicking this button you will be returned to the Equation