5: Maintenance
5.4 Factory Calibration
(yearly)
Certifier
®
FA Test System flow modules are designed for one year of
normal use following each factory calibration.
If the test system has been dropped or liquid has penetrated any of its
components, do not use, and return to the factory for calibration.
Recalibrated flow modules come with a certificate of calibration and a
summary of performance before and after the calibration. A factory
calibration consists of pressure transducer calibration over the full
range of pressures and calibration over the full range of flows. All
calibration datum are stored in the flow modules, so the controller
module does
not
require calibration. Therefore, it is not necessary to
return the controller module for factory calibration.
Follow the steps in
Section 5.5
to return Certifier
®
FA Test System
flow modules for factory calibration.
5.5 Return Procedure
Follow these steps to return Certifier
®
FA Test System flow modules
for factory calibration:
1. Contact one of the following offices to make service
arrangements and obtain a Return Material Authorization (RMA)
number or use our online RMA form at www.tsi.com.
2. Package the flow modules carefully to avoid damage during
shipping.
NOTE:
It is
not
necessary to return the controller module for
factory calibration.
U.S. & International
United Kingdom
TSI Incorporated
TSI Instruments Ltd.
500 Cardigan Road
Tel:
(44) 1494 459200
Shoreview MN 55126-3996
Fax:
(44) 1494 459700
USA
E-mail:
Tel:
(800) 874-2811 /
Website:
www.tsiinc.co.uk
+1(651) 490-2811
Fax
:
+1(651) 490-3824
Germany
E-mail:
TSI GmbH
Website
:
www.tsi.com
Tel:
+49 241-52303-0
Fax
:
+49 241-52303-49
E-mail:
Website:
www.tsiinc.de