31
6 Specifications
NOTE:
Specifications are subject to change without notice.
6.1 Physical
Dimensions
Controller module: 13.2 cm x 7 cm x 3.3 cm (5.2 in. x 2.8 in. x 1.3 in.).
High Flow module: 18 cm x 6.6 cm x 4 cm (7.1 in. x 2.6 in. x 1.6 in.).
Low Flow module: 12.7 cm x 5.1 cm x 2.8 cm (5.0 in. x 2.0 in. x 1.1 in.).
Flow
connectors
High Flow module:
flow inlet: 22-mm female ISO taper.
flow outlet: 22-mm male ISO taper.
Low Flow module:
flow inlet: 0.25-in. barb.
flow outlet: 0.25-in. barb.
Weight
Approximately 1.36 kg (3 lb) for the standard kits.
6.2 Environmental
Temperature
Operating: 5 to 40
C (41 to 104
F). 15 to 80% relative humidity from
5 to 31
C decreasing linearly to 15 to 50% relative humidity at 40
C.
Storage: -40 to 70
C (-40 to 158
F) at 10 to 95% relative humidity.
Atmospheric
Pressure
Operating: 57.1 to 106 kPa (8.28 to 15.37 psia).
Storage: 50 to 106 kPa (7.25 to 15.37 psia).
Conditions
Indoor Use
Operating Altitude up to 4000 m (13,000 ft)
Pollution degree I or II
6.3 Power
Battery life
15 to 20 hours.
Battery type
Four commercially available AA batteries.
6.4 Test Measurements
(see notes at end of section)
Measurement
High Flow Module
Low Flow Module
Flow and Peak Flow
Range
0 to 300.0 SLPM.
0.01 to 15.00 SLPM.
Accuracy
Air and oxygen:
2% of reading
or
0.075 SLPM, whichever is
greater.
Air/oxygen mixtures:
4% of
reading or
0.1 SLPM, whichever
is greater.
Air and oxygen:
2% of
reading or
0.010 SLPM,
whichever is greater.
Nitrous oxide:
4% of reading
or
0.025 SLPM, whichever is
greater.