A-1
A P P E N D I X A
S p e c i f i c a t i o n s
All specifications meet or exceed JIS B9921. They are subject to
change without notice.
Size Range
0.3 to 25 µm
Channel Sizes
0.3 µm and 5.0 µm fixed; 0.5, 1.0, 2.0, or 2.5 µm
selectable middle channel.
Counting Efficiency
50% at 0.3 µm; 100% for particles > 0.45 µm
(per ISO 21501-4 and JIS)
Concentration Limits
2,000,000 particles/ft
3
at 5% coincidence loss
Light Source
Long life laser diode
Zero Count Level
<1 count / 5 minutes (per ISO 21501-4 and
JIS B9921)
Flow Rate
0.1 CFM (2.83 L/min)
Calibration
NIST traceable with TSI calibration system
Sample Probe/Tubing
Isokinetic sampling probe
Sampling Modes
Manual and automatic
Sampling Time
1 second to 99 minutes 59 seconds
Sampling Frequency
1 to 999 cycles or continuous
Exhaust
Internal HEPA filter
Vacuum Source
Internal pump
Communication Mode
USB serial output
Data Storage
1,500 sample records
Status Indicators
Low battery and service
Display
3.2-in. (8.1 cm) 160 x 160 monochrome LCD
External Surface
High impact injection molded plastic
AC Power (power to AC
adapter)
110 to 240 VAC 50-60 Hz Universal in-line power
supply with country specific plugs
DC Power (power to
instrument)
9 VDC @ 1.5 A
Battery
Rechargeable Li-Ion
Battery Life
Up to 4.5 hours of continuous use
Recharge Time
< 3 hours
Dimensions (L x W x H)
9.1 x 6.3 x 2.4 in. (23 x 11 x 6.2 cm) (without
isokinetic inlet)
Weight
1.3 lbs (0.58 kg) with battery
Standards
CE, JIS B9921, ISO 21501-4
Warranty
1 year (extended warranties available)
Summary of Contents for AEROTRAK 9303
Page 2: ...ii ...
Page 6: ......
Page 16: ...1 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 32: ...3 12 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 34: ...4 2 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 36: ...5 2 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 42: ...7 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 45: ...Specifications A 3 D i m e n s i o n a l D i a g r a m ...
Page 46: ...A 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 49: ...Index 3 weight A 1 Z zero check maintenance 5 1 performing 3 2 zero count level A 1 ...
Page 50: ......