5-1
C H A P T E R 5
M a i n t e n a n c e
N o t e
There are no user-serviceable parts inside this instrument. Opening the
instrument case may void the warranty. TSI recommends that you return
the A
ERO
T
RAK
™
Airborne Particle Counter to the factory for any required
maintenance or service not described in this manual.
M a i n t e n a n c e S c h e d u l e
TSI recommends annual factory cleaning and calibration for the
A
ERO
T
RAK
™
Airborne Particle Counter. See
Recommended Field Maintenance Schedule
Item
Frequency
Zero check
Daily (or as defined by application).
Factory cleaning and calibration
Annually.
Cleaning the instrument enclosure
As needed
Z e r o C h e c k
The zero check ensures that the instrument is properly assembled
and free from leaks, residual particles and electronic noise. Please
see Chapter 3, “
” for detailed instructions on performing
the zero check.
C l e a n i n g t h e I n s t r u m e n t E n c l o s u r e
To clean the enclosure, dampen a lint-free cloth and gently wipe
the surface until surface contamination is removed.
Summary of Contents for AEROTRAK 9303
Page 2: ...ii ...
Page 6: ......
Page 16: ...1 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 32: ...3 12 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 34: ...4 2 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 36: ...5 2 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 42: ...7 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 45: ...Specifications A 3 D i m e n s i o n a l D i a g r a m ...
Page 46: ...A 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 49: ...Index 3 weight A 1 Z zero check maintenance 5 1 performing 3 2 zero count level A 1 ...
Page 50: ......