![TSI Incorporated AEROTRAK 9303 Operation Manual Download Page 30](http://html1.mh-extra.com/html/tsi-incorporated/aerotrak-9303/aerotrak-9303_operation-manual_1178058030.webp)
3-10
A
ERO
T
RAK
™ Handheld Airborne Particle Counter
BUFFER RECORD
Record#:
0313
Date:
12/27/08
Time:
05:50:45
0.3µm
000566473
0.5µm
000534571
5.0µm
000072056
Location:
111
Sample Time:
01:00
Hold Time:
00:00
<
>
Example of a Buffer Record
Use the < and > softkeys to scroll through the records, Enter to go
back one screen and
ESC
to go back to the Buffer Menu.
N o t e
If the buffer is filled to its capacity of 1500 records, the instrument will
continue to count and save records but the next record after 1500 will
be saved in record number 1, then 2, etc. This way the most recent
1500 records are always preserved.
U t i l i t y M e n u
The Utility Menu lets you adjust the pump speed, turn data
logging on and off, and select the sampling location.
UTILITY MENU
Pump Speed
099
Data Logging ON
Location
120
Utility Menu
The following table describes the options of this menu and the
parameters you can set.
Summary of Contents for AEROTRAK 9303
Page 2: ...ii ...
Page 6: ......
Page 16: ...1 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 32: ...3 12 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 34: ...4 2 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 36: ...5 2 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 42: ...7 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 45: ...Specifications A 3 D i m e n s i o n a l D i a g r a m ...
Page 46: ...A 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 49: ...Index 3 weight A 1 Z zero check maintenance 5 1 performing 3 2 zero count level A 1 ...
Page 50: ......