S a m p l e M e n u
The Sample Menu sets sample mode, cycle mode, sample time,
hold time, cycle count, sample units, and channel 2 size.
SAMPLE MENU
Sample Mode
COUNTS
Cycle Mode
AUTO
Sample Time
01:00
Hold Time
00:30
Cycle Count
010
Sample Units
LITER
Chan 2 Size
0.5µm
<
>
Sample Menu
The following table describes the options of this menu and the
parameters you can set.
Option
Description
Parameters
Sample Mode
Use
<
and
>
to set the
sampling mode.
Concentration (see note
below); counts
Cycle Mode
Use
<
and
>
to set the
sampling mode.
Manual – the counter will start
when the RUN soft key is
pressed and stop only when the
STOP soft key is pressed.
Auto – the counter will start
when the RUN soft key is
pressed and turn on and off
automatically to follow the
sample time, hold time and
cycle count parameters below.
Sample Time
Press the
Enter
key to go to
a secondary screen where
you can set the sample
time.
Minutes and seconds up to
99:59 (see note below)
Hold Time
Press the
Enter
key to go to
a secondary screen where
you can set the time
between samples.
Minutes and seconds up to
99:59
Cycle Count
Press the
Enter
key to go to
a secondary screen where
you can set the number of
samples to take.
1 to 999
Sample Units
Use
<
and
>
to set the
sampling units.
Liter; cubic feet
Summary of Contents for AEROTRAK 9303
Page 2: ...ii ...
Page 6: ......
Page 16: ...1 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 32: ...3 12 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 34: ...4 2 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 36: ...5 2 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 42: ...7 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 45: ...Specifications A 3 D i m e n s i o n a l D i a g r a m ...
Page 46: ...A 4 AEROTRAK Handheld Airborne Particle Counter This page intentionally left blank ...
Page 49: ...Index 3 weight A 1 Z zero check maintenance 5 1 performing 3 2 zero count level A 1 ...
Page 50: ......