DRIFT TEST
If the gauge passes the stat test, but shows long-term drift between
measurements, perform a
drift test
to check the long-term drift. A
drift test consists of five counts taken 3–8 hours after a stat test. The
gauge sets the pass/fail limits based on the percent difference
between the average of the stat test and drift test results. If the
percent difference exceeds 0.5 percent for density counts or 1
percent for moisture counts, the drift test
fails
. If the drift test fails,
ensure that the gauge setup and testing were conducted correctly,
and take an additional test. If the second test fails, call the nearest
Troxler service center for assistance.
NOTE
The drift test consists of five 4-minute count, whereas
the stat test consists of twenty 1-minute counts.
Therefore, each test takes approximately twenty minutes
to conduct.
DO NOT turn the gauge off between a stat test and a
drift test. The stat test must be current. The gauge must
not be moved between the stat and drift tests to
eliminate possible failure due to positioning changes.
To access the
Drift Test
functions, press the
〈
Drift Test
〉
button on
the
Diagnostics Menu
(see Figure C–1). The PDA displays the
Drift Test Menu
shown in Figure C–6.
Taking a Drift Test
With the gauge still in an area free of other nuclear gauges or
radioactive sources and on the reference standard block, press
〈
Take Drift Test
〉
. The PDA displays the date and time of the last
stat test and asks whether to continue with the drift test.
NOTE
If it has been less than four hours or more than eight
hours since the last stat test, then the gauge allows the
operator to exit the drift test. To exit and take another
stat test, press
〈
No
〉
at the prompt. To continue with the
drift test, press
〈
Yes
〉
at the prompt.
Model 3451
Appendix C–5
C.
T
RO
UBL
E
S
HO
O
T
ING
Summary of Contents for Enhanced RoadReader Plus 3451
Page 7: ...Model 3451 vii NOTES...
Page 17: ...xvi NOTES...
Page 29: ...1 12 NOTES...
Page 33: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 85: ...4 22 NOTES...
Page 106: ...Model 3451 6 3 Figure 6 1 Offset Menu 6 OFFSETS CALIBRATION...
Page 114: ...Model 3451 6 11 Figure 6 10 Moisture Offset Erase Prompt 6 OFFSETS CALIBRATION...
Page 131: ...6 NOTES...
Page 141: ...7 10 Figure 7 5 Project Data Spreadsheet...
Page 145: ...7 14 Figure 7 8 Project Setup Display Figure 7 9 Standard Counts Display...
Page 147: ...7 16 Figure 7 10 Project Status Display Figure 7 11 Source Decay Display...
Page 149: ...7 18 Figure 7 12 Stat Test Display Figure 7 13 Drift Test Display...
Page 153: ...7 22 NOTES...
Page 166: ...Model 3451 Appendix A 13 Figure A 3 Cleaning the Sliding and Fixed Blocks...
Page 174: ...Model 3451 Appendix A 21 NOTES...
Page 175: ......
Page 183: ...Appendix B 8 NOTES...
Page 221: ...Appendix D 6 NOTES...
Page 223: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 225: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 227: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 229: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 235: ...Appendix G 4 NOTES...