Appendix D–4
Type A Package Testing Results for Model 3451
Water Spray Test:
Subject the package to a water spray simulating rainfall
of approximately 2 inches per hour for 1 continuous
hour.
Results:
Engineering evaluation – Water exposure of the
magnitude required by the regulations would not affect
the shielding or containment integrity of the package.
Vibration Test:
The package was vibrated with a displacement of 0.1"
at approximately 12 Hz for a period of 24 continuous
hours.
Results:
No physical damage to the package was observed.
Free Drop Test:
The package was dropped from a height of 4 feet onto a
non-yielding surface from a position to cause
maximum damage to the package.
Results:
The case was scratched due to the abrasiveness of the
concrete, but no other physical damage to the package
was observed.
Penetration:
The package was placed on a non-yielding surface. A
1-1/4-inch diameter, 13-pound steel cylinder with a
hemispherical end was dropped in the vertical position
from a height of 40 inches onto the package to a point
to cause maximum damage to the package.
Results:
No physical damage to the package was observed.
Compression:
Package was placed on a non-yielding surface and
subjected to a compressive load of at least 13
kilopascals multiplied by the vertically projected area
of the package, in square feet, for 24 continuous hours.
Results:
No physical damage to the package was observed.
Weight of package as tested: 96 pounds
Date test completed: July 1996
Figure D–1. Type A Package Testing Results
Summary of Contents for Enhanced RoadReader Plus 3451
Page 7: ...Model 3451 vii NOTES...
Page 17: ...xvi NOTES...
Page 29: ...1 12 NOTES...
Page 33: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 85: ...4 22 NOTES...
Page 106: ...Model 3451 6 3 Figure 6 1 Offset Menu 6 OFFSETS CALIBRATION...
Page 114: ...Model 3451 6 11 Figure 6 10 Moisture Offset Erase Prompt 6 OFFSETS CALIBRATION...
Page 131: ...6 NOTES...
Page 141: ...7 10 Figure 7 5 Project Data Spreadsheet...
Page 145: ...7 14 Figure 7 8 Project Setup Display Figure 7 9 Standard Counts Display...
Page 147: ...7 16 Figure 7 10 Project Status Display Figure 7 11 Source Decay Display...
Page 149: ...7 18 Figure 7 12 Stat Test Display Figure 7 13 Drift Test Display...
Page 153: ...7 22 NOTES...
Page 166: ...Model 3451 Appendix A 13 Figure A 3 Cleaning the Sliding and Fixed Blocks...
Page 174: ...Model 3451 Appendix A 21 NOTES...
Page 175: ......
Page 183: ...Appendix B 8 NOTES...
Page 221: ...Appendix D 6 NOTES...
Page 223: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 225: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 227: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 229: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 235: ...Appendix G 4 NOTES...