![Troxler Enhanced RoadReader Plus 3451 Manual Of Operation And Instruction Download Page 130](http://html1.mh-extra.com/html/troxler/enhanced-roadreader-plus-3451/enhanced-roadreader-plus-3451_manual-of-operation-and-instruction_1172542130.webp)
Model 3451
6–27
Number of Averages
The
Number of Average
s function allows averaging of up to 20
readings from a single test location when testing thin layer asphalt.
This function is used only when the gauge is measuring in thin layer
mode. When the gauge is using this function, the PDA
automatically prompts for repositioning of the gauge from a test
location and begins another measurement. After all measurements
have been completed, the gauge calculates the average density of
the test material. When taking counts on thin layer asphalt, Troxler
recommends averaging four one-minute readings (see the
To access this function, press the
〈
Number of Averages
〉
button
on the
Thin Lift Special Calibration
menu (see Figure 6–21).
The PDA requests the number of thin layer averages, as shown in
Figure 6–28. Enter the desired number of readings (1 through 20)
for the gauge to average.
Press
〈
ok
〉
to return to the
Thin Lift Special Calibration
menu.
Figure 6–28. Input Number of Averages
6.
O
F
F
S
E
T
S
&
C
A
L
IB
R
AT
IO
N
Summary of Contents for Enhanced RoadReader Plus 3451
Page 7: ...Model 3451 vii NOTES...
Page 17: ...xvi NOTES...
Page 29: ...1 12 NOTES...
Page 33: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 85: ...4 22 NOTES...
Page 106: ...Model 3451 6 3 Figure 6 1 Offset Menu 6 OFFSETS CALIBRATION...
Page 114: ...Model 3451 6 11 Figure 6 10 Moisture Offset Erase Prompt 6 OFFSETS CALIBRATION...
Page 131: ...6 NOTES...
Page 141: ...7 10 Figure 7 5 Project Data Spreadsheet...
Page 145: ...7 14 Figure 7 8 Project Setup Display Figure 7 9 Standard Counts Display...
Page 147: ...7 16 Figure 7 10 Project Status Display Figure 7 11 Source Decay Display...
Page 149: ...7 18 Figure 7 12 Stat Test Display Figure 7 13 Drift Test Display...
Page 153: ...7 22 NOTES...
Page 166: ...Model 3451 Appendix A 13 Figure A 3 Cleaning the Sliding and Fixed Blocks...
Page 174: ...Model 3451 Appendix A 21 NOTES...
Page 175: ......
Page 183: ...Appendix B 8 NOTES...
Page 221: ...Appendix D 6 NOTES...
Page 223: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 225: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 227: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 229: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 235: ...Appendix G 4 NOTES...